LB 470RID Level
9 Troubleshooting
56925-1BA2 Rev.02, 05/2021
185
9.2.6
Process Connection
Code
Message
NAMUR107 Help Text
M701
Current output
F
Deviation of analog output value from
feeback value is too high. Calibrate ana-
log output. Contact Berthold service, if
this event occurs repeatedly.
M702
Current loop open
F
Current output loop open. Check cable
connection.
M703
Software update
C
Indicates that a software update is run-
ning. No action necessary. The device
automatically returns to measuring
mode after the software update is fin-
ished.
M799
Internal program error
F
Internal software failure. Restart the de-
vice. Contact Berthold service, if this
event occurs repeatedly.