9 Troubleshooting
LB 470RID Level
180
56925-1BA2 Rev.02, 05/2021
9.2.2
Application
Code
Message
NAMUR107 Help Text
M301
Default parameter set
M
Device not calibrated. Measurement
with default parameters. Calibrate de-
vice
M302
Decay compensation
S
Decay compensation failed. Contact
Berthold service, if this event occurs re-
peatedly.
M303
Det Temp LL failure
F
Detector temperature at / below lower
limit. Limit value can be configured. Fac-
tory setting: -20°C.
M304
Det Temp UL failure
F
Detector temperature at / above upper
limit. Limit value can be configured. Fac-
tory setting: 60°C
M305
Det Temp LL OOS
S
Detector temperature at / below lower
limit. Limit value can be configured. Fac-
tory setting: -20°C.
M306
Det Temp UL OOS
S
Detector temperature at / above upper
limit. Limit value can be configured. Fac-
tory setting: 60°C.
M307
Stray radiation
S
Interference radiation detected. Meas-
urement stopped.
M308
Source replacement
M
Source replacement date reached. Re-
place source.
M309
Application stopped
C
Measurement stopped
M310
PV calc not possible
S
Process value could not be calculated.
Check measuring range and calibration.
M311
Backup process
C
Backup in process.
M312
Restore process
C
Restore in process.
M320
PV LL failure
F
Process value at / below lower limit.
M321
PV UL failure
F
Process value at / above upper limit.
M322
PV LL OOS
S
Process value at / below lower limit.
M323
PV UL OOS
S
Process value at / above upper limit.
M324
Level under 0%
S
Level below 0%. Check measuring range
and calibration.
M325
Level over 100%
S
Level above 100%. Check measuring
range and calibration.
M326
GPC out of spec
S
Compensation factor of Gas Properties
Compensation has reached its limit.
Check process.
M327
No GPC detector
S
No detector for Gas Properties Compen-
sation found. Connect / configure
detector.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...