8.5
Network Data Log
Diagnostics | Network Data Log
In the window “Network data log”, the transmission of log data via the ethernet
network can be started. With a log program, the data can be displayed on the PC.
IMPORTANT
The PC and the LB 47x have to be in the same IP subnet.
Observe the notes in chapter 7.3.1 – Network.
The network logger utilizes the Telnet protocol to send data over the local net-
work. There are terminal emulator programs that can read this data and save it
in a log file. Note the following links:
https://support.microsoft.com/help/2801292
1
2
3
4
Button
<Start>
Button
<Stop>
Display field / Input field „Logging Interval“
Display field „Logging State“
Fig. 143 Network Data Log (started)
1.
Click Stop if necessary. Click on the entry field “Logging interval” (Fig. 144,
item 3) and enter an interval.
2.
Start the network data log with the button
<Start>
(Fig. 144, item 1).
3.
Make a note of the EVUs IP address (
Device Settings | Settings | System | Net-
work
).
4.
Enter the following input in the command line of the PC:
> telnet "IP Address" -f "LogFileName" (for example 192.168.83 -f C:\\Us-
ers\Test\log.txt).
The log data is displayed and saved in the created log file.
Use the key combination "CTRL" and "+" to stop the recording of the log data.
Summary of Contents for DUO XPERT LB 470RID
Page 2: ......
Page 12: ...1 About this Operating Manual LB 470RID Level 12 56925 1BA2 Rev 02 05 2021...
Page 16: ...2 Safety LB 470RID Level 16 56925 1BA2 Rev 02 05 2021...
Page 34: ...3 System Description LB 470RID Level 34 56925 1BA2 Rev 02 05 2021...
Page 42: ...4 Installation LB 470RID Level 42 56925 1BA2 Rev 02 05 2021...
Page 207: ...Level F llstand Technical Information Technische Information 56925TI1L Rev No 04 05 2021...