Appendix
LB 983 NightSHADE II Operating Manual
120
10.1.4 Temperature Controlled Measurement Table
The temperature controlled measurement table is used for measurements which require a tem-
perature controlled environment.
If you are working with the heated macro table, plug the connector into the 9-pole Sub-D-
connector on the power box at the center top inside the darkbox (see 5.3.3.) The device must
be switched off, since after the NightSHADE is switched on the heating system is detected and
may get initialized.
If the heating system is plugged in during ongoing operation, it will work only after the device
has booted up again. This is done by clicking on .
and selecting
<Boot Device>
from
the
<Reference>
dropdown list.
After approx. 5 to 10 minutes the heated macro measurement table has reached
its working temperature (preset 37°C). The current temperature is displayed in
the bottom status bar of the indiGO user interface
.
The default temperature can be defined in the configuration backend of the
software accessible through the instrument controller. Right click on the control-
ler icon and log-in with a valid user account information selecting Setup from the
context menu.
In the
<Presets> <Heater>
section the default temperature can be set. When the checkbox at
the bottom is checked indiGO will set the default temperature at system initialization.
Summary of Contents for 55393 BA2
Page 4: ...Prefatory Comments LB 983 NightSHADE II Operating Manual 4 ...
Page 46: ...Quick Start LB 983 NightSHADE II Operating Manual 46 tems ...
Page 80: ......
Page 101: ...LB 983 NightSHADE Operating Manual indiGO Software Documentation 101 ...
Page 104: ...indiGO Software Documentation LB 983 NightSHADE II Operating Manual 104 ...