Shields for Flange Mounting
6 Usage
44192BA2, Rev.03, 07/2021
31
j
WARNING
Danger of injury by falling loads!
Assemble and disassemble the shield head with a crane and exclusively use
the lifting eyes for attaching the sling gear.
1.
Remove the padlock (Fig. 13, item 1) on the cover / shield head (Fig. 13,
item 2).
2.
If necessary, remove the screws at the cover / shield head.
3.
Lift the shield head vertically upwards or remove the cover by pulling it
upwards.
The coupling element of the source holder and the clamping screw
becomes accessible.
NOTICE
The sources are secured in the delivery state with clamping screws in the
shield.
With the shutter open, open the clamping screw only when every
source is attached to the steel rope or to the flexible shaft!
1
2
3
4
Coupling element of the source holder
Coupling element for steel rope or flexible shaft
Steel rope
Flexible shaft
Fig. 14
Latching the steel rope or the flexible shaft into the source holder