Chapter 5 Explanations
58
Chapter 5.
Explanations
5.1
Zero Count Rate
The zero count rate is the count rate caused by natural environ-
mental radiation. In contrast to radiation coming from the source,
the zero count rate remains constant. The zero count rate is large-
ly dependent on the scintillator volume. An error in measuring the
zero count rate may result in switching errors.
If no zero count rate values have been entered or read-in, the de-
vice works with the standard value set in the factory. The standard
value is an approximate value defined by the current detector
code. Since background radiation is dependent on the geographical
location, this value should be read in as soon as possible.
The zero count rate is needed for the correct calculation of the
decay compensation.
When reading in the zero count rate, please keep in mind that "re-
sidual radiation", even from a closed shielding container, may dis-
tort the empty count rate value. Gross falsification may lead to
false alarms being triggered over a longer period of operation
(months/years).
5.1.1
Determining the Zero Count Rate
The zero count rate is determined according to the best method.
Often, however, the best method cannot be employed, depending
on the conditions during production or start-up. Below, we will de-
scribe three methods for determining the zero count rate.
a) Best method
With empty or full container, but
without
source.
Figure 6:
Determining the
zero count rate
with no source
installed
with empty container
with full container
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...