Chapter 3 Parameters
42
Detector high voltage
Show or set current high voltage of the detector
(only FSK detectors)
Input
HV mode
-1
Automatic
500 -1300
Fixed value
The automatic HV-control keeps the detector free from drifts in
case of temperature fluctuations and ageing.
Enter a value between 500 and 1300 V to set the detector to man-
ual high voltage.
Note!
To ensure a reliable function of the measurement, the HV has to
be set to automatic.
If the count rate of a NaI detector drops below 100 counts, the HV
is frozen. If it rises again above 200 counts, the HV is again con-
trolled automatically.
HV start value
Set start value of high voltage control
(only FSK detector)
Input
HV start value
0
Disabled
500 - 1300
Enabled
A HV start value fulfills the following tasks:
Default value
After turning on power, the HV starts with this value. The
measurement will be ready for operation quickly.
Limit value
for the HV control range.
It limits the HV control range of the detector to max. +40%
and min. 20% of the HV start value. A HV that is too high and
therefore detrimental to the photomultiplier cannot be generat-
ed. An error message is output if the HV touches one of the
cut-off regions (see page 71).
The start value is determined in the factory and stored in the de-
tector. If the photomultiplier has to be replaced, the automatic HV
value of code 53 automatically has to be entered in code 54.
C o d e 5 3
C o d e 5 4
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...