Chapter 3 Parameters
37
Bulk cone diameter (mm)
Enter diameter of the bulk cone where switching is to be
performed.
You have to enter the bulk cone di-
ameter "d" in mm if measurements
on bulk goods should trigger an
alarm at a certain bulk cone diame-
ter.
Then the limit value indicates an
alarm as soon as the bulk cone has
reached the specified diameter "d".
Provided the value in code 38 >0, the switch point in code 17/18 is
then calculated based on the bulk density of code 35, the nuclide
defined in code 5 and the bulk cone diameter of code 38.
If a gas density has been entered in code 35, this will also be tak-
en into account in calculating the switch point.
For the entries to take effect, you have to calibrate with code 36.
If the isotope is changed later, the switch point will be adjusted
automatically, and you do not have to enable code 36.
With 0 you can turn this function off.
Half-value layers
Enter product absorption at full container
Using the number of half-value layers (HVL) the full count rate can
automatically be calculated based on the empty count rate.
For code 39 to be effective in the professional mode, code 31 has
to be set to -1 and code 33 and 35 to 0.
The number of HVL’s is dependent on the measuring path through
the product being measured and the bulk density (with liquids:
density) of the product. In most cases, the standard value 2 HVL is
adequate for calculating the full count rate automatically.
The HVL can be determined as follows:
HVL = rho * d / k
rho
: density of product being measured (g/cm³)
d
:
measuring path = internal container diameter (mm)
k
Co-60: 157
k
Cs-137: 110
C o d e 3 8
Bulk cone measurement
C o d e 3 9
Formula
d (mm)
Cs-137
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...