Chapter 3 Parameters
36
Note!
If the parameters
- time constant (code 12)
- switching threshold (code 17/18)
- hysteresis (code 19)
have been set to automatic, there is only a very small risk that
false values have been set.
We recommend to enable code 36 after every parameter change to
ensure that an incorrect setting will be recognized immediately.
Calibration error - Table
Error
code
Sub-
code
Designation
Explanation
Remedy
35
01
ECR < FCR
Empty count rate
lower than zero count
rate.
Determine empty
count rate or / and
zero count rate
new.
02
ECR < FCR
Empty count rate
lower than full count
rate.
Determine empty
count rate or / and
full count rate new.
03
FCR < ZCR
Full count rate lower
than zero count rate.
Determine full count
rate or / and zero
count rate new.
05
Distance
switch point
to FCR
Distance switch point
to full count rate is
too small. Danger of
switching errors.
Adjust switch point,
or increase time
constant.
06
Distance
switch point
to ECR
Distance switch point
to full count rate is
too small. Danger of
switching errors.
Adjust switch point,
or increase time
constant.
07
Hysteresis
too large
Hysteresis too large.
Increase time
constant or reduce
hysteresis.
08
Time constant
too small
Time constant too
small
Increase time con-
stant or set switch
point further into
the center.
Counting time for calibration (in seconds)
Enter length of time for count rate averaging
(Only in the professional mode)
Here you can enter the length of time to be used for averaging of
the calibrated count rates in code 30, 31 and 32. The factory set-
ting is 60 seconds.
Input limits: 5 … 600s.
Table 4:
Calibration error
C o d e 3 7
Summary of Contents for LB4710-050
Page 2: ......
Page 4: ......
Page 8: ......
Page 32: ......
Page 33: ...Chapter 3 Functional Safety Mini Switch LB 471 33...
Page 34: ...Chapter 3 Functional Safety Mini Switch LB 471 34...
Page 35: ...Chapter 3 Functional Safety Mini Switch LB 471 35...
Page 44: ......
Page 58: ......
Page 98: ......
Page 104: ......
Page 110: ...Chapter 14 Certificates Mini Switch LB 471 110...
Page 111: ...Chapter 14 Certificates Mini Switch LB 471 111...
Page 112: ...Chapter 14 Certificates Mini Switch LB 471 112...
Page 113: ...Chapter 14 Certificates Mini Switch LB 471 113 14 2 ATEX Certificate NaI Detector...
Page 114: ...Chapter 14 Certificates Mini Switch LB 471 114...
Page 115: ...Chapter 14 Certificates Mini Switch LB 471 115...
Page 116: ...Chapter 14 Certificates Mini Switch LB 471 116...
Page 117: ...Chapter 14 Certificates Mini Switch LB 471 117 14 3 ATEX Certificate GM Detector...
Page 118: ...Chapter 14 Certificates Mini Switch LB 471 118...
Page 119: ...Chapter 14 Certificates Mini Switch LB 471 119 14 4 EC Declaration of Conformity...
Page 120: ......
Page 141: ...Mini Switch LB 471 141 Notes...
Page 142: ...Mini Switch LB 471 142 Notes...
Page 145: ......
Page 147: ......
Page 149: ......
Page 151: ......
Page 161: ......
Page 195: ......
Page 215: ......
Page 219: ......
Page 223: ...Mini Switch LB 471 78 Notes...
Page 224: ...Mini Switch LB 471 79 Notes...
Page 227: ...Mini Switch LB 471 82 Y Year 21 Z Zero count rate 8 30 32 58...