Chapter 6 Instructions for use: Ikus
6.2.8 Browsing in the message window
INSTRUCTION
1.
Press the
<7>
key to move the cursor in the message window.
2.
Scroll through the messages: either 1 message at a time with
<
↓
>
/
<
↑
>
; or 4
messages at a time with
<Bild-
↓
>
/
<Bild-
↑
>
.
3.
Exit the message window by pressing either
<Esc>
or
<Enter>
. The field
Log
off
is automatically activated and appears with a black background.
6.3
Stopping the blood pump(s) and switching off the Ikus
Fig. 6-17
Drive pause (with stop options)
The pop-up menu
Drive pause
offers the following options:
•
Drive pause
•
Pause left
•
Pause right
(only in biventricular operation)
•
Drive OFF
6.3.1 Drive pause: stopping the Ikus temporarily
IMPORTANT: When restarting the system from a separate mode there can be an
unintentional switch to the synchronous mode, although
L/R separate
continues to be
displayed. The pressure graphs, however, distinctly indicate the synchronous mode.
INSTRUCTION
1.
Select
Drive pause
, then press
<Enter>
to confirm. Respond to the prompt in
the dialog window by pressing the
<X>
key or the
<1>
key. The Ikus will stop.
The view
Select operating mode
is displayed.
Fig. 6-18
Select operating mode
2.
Select
Univentricular (UVAD)
or
Biventricular (BVAD)
, then confirm the
selected operating mode with
<Enter>
. In biventricular mode the view
Pump
size and single-step mode
appears. In univentricular mode, the seal test is
performed first and then the view
Pump size and single-step mode
appears.
1000721x09 Revision 8
77
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...