Chapter 5 Description: Ikus
5
Description: Ikus
5.1
Overview
The electro-pneumatic Ikus generates the suction and driving pressures required to
drive the blood pump(s). The driving unit contains the pneumatic and electronic
components as well as a laptop computer which serves as an interface to the operator.
7
1
6
1
laptop computer with monitor program
2
lifting bar
3
mains cable
5
4
driving tubes
5
connection panel
6
handle
7
display and operating panel
2
4
3
Fig. 5-1
Ikus
Pneumatic systems
The Ikus has 3 pneumatic systems operating independently of each other. One
pneumatic system is required for each blood pump, while the 3
rd
serves as an
emergency backup. Each pneumatic system includes:
•
a compressor
• pressure and suction limiters
• pressure and vacuum cylinders
• control electronic
• control valves
The compressor, the pressure and the suction limiters deliver constant pressure
conditions in the pressure and vacuum cylinders. The control valves at the outlet of
each cylinder allow optimal adjustment of the positive and negative (suction) pressure
values.
Control computer
The Ikus system has 2 control computers operating independently of each other: the
active (main) control computer and the backup control computer.
Laptop with monitor program
Messages and pressure graphs in the monitor program inform the user of the current
status and working condition of the system. In addition, the laptop computer is used
1000721x09 Revision 8
57
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...