Chapter 17 EMC tables
17 EMC tables
17.1 Essential Performance
The following essential performance was verified in the electromagnetic immunity
tests: The driving unit must drive the EXCOR blood pumps with the set parameters.
This was controlled by monitoring the following acceptance criteria:
•
The pump rate may not deviate by more than 10% or 50 ms (the higher value is
valid).
•
The relative systolic duration may not deviate by more than 10% or 50 ms (the
higher value is valid).
•
The driving pressure in engaged condition at the end of the systole or diastole
may not deviate by more than 10% or 20 mmHg (the higher value is valid).
•
No failure alarms may occur.
•
No undesired change may occur between mains and battery operation.
•
In battery operation the battery alarms must occur correctly.
•
No switch to emergency pulse mode may occur.
17.2 Electromagnetic emissions
The Ikus is intended for use in the electromagnetic environment specified below. The
customer or the user of the Ikus should ensure that Ikus is used in such an
environment
Emissions test
Compliance
Electromagnetic environment -
guidance
RF emissions
CISPR 11
Group 1
The Ikus uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
RF emissions
CISPR 11
Class B
The Ikus is suitable for use in all
establishments, including domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Tab. 17-1
Emissions characteristics
17.3 Electromagnetic immunity - part 1
The Ikus is intended for use in the electromagnetic environment specified below. The
customer or the user of the Ikus should ensure that it is used in such an environment.
1000721x09 Revision 8
231
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...