Chapter 1 Important safety information
Cleaning the cannulae and transcutaneous exit site: Do not use any
acetone or petroleum based products near the cannulae and the
transcutaneous exit site.
We recommend using chlorhexidine to clean the cannulae and
transcutaneous exit site.
IMPORTANT: Do not use any corrosive or colored solutions or organic
solvents to clean the cannulae and transcutaneous exit site as they
may alter the surface of the product.
Prevent liquids from spilling into the Ikus. Otherwise there is a risk of an
electric shock and of a malfunction of the Ikus.
1.1.10 Maintenance
WARNING
If the Ikus is not in operation, it requires maintenance every 6 months.
If it is in operation, it requires maintenance after every 2000 hours of
use. (In case of continuous operation, approx. 3 months)
If the ambient temperature is continuously above +30°C during
operation, the maintenance interval or life of the batteries is reduced.
The Ikus shall only be serviced by Berlin Heart GmbH/ Berlin Heart, Inc.
or those authorized by Berlin Heart GmbH. For this reason, this
document does not contain any circuit or wiring diagrams. The
maintenance is performed based upon the
Maintenance Instructions
.
This document also includes a technical description of the Ikus.
Only replacement parts approved by the manufacturer may be used for
repairs and servicing. Otherwise there is a risk of functional limitation
or permanent damage of the Ikus.
1.1.11 Errors and corrective measures
WARNING
Any time an error message has occurred, visually check that the blood
pump(s) is (are) filling and ejecting completely over a period of several
pump cycles, then address the error message with the appropriate
corrective action.
Check all information and messages in the message window of the
monitor program at least every 4 hours. Take the necessary measures
and (if required) notify the service department of Berlin Heart, Inc. The
message window only shows a limited number of messages. Otherwise
there is a risk that older information and messages can no longer be
read and therefore corrective actions for older messages may no
longer be possible.
1000721x09 Revision 8
14
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...