Chapter 14 Error Messages and corrective measures
14.8 System 3 (backup) is defective!
System 3 is defective!
The backup pneumatic system is detected as a fault during backup operation.
The Ikus runs in emergency operating mode. Provide the patient immediately with a
replacement Ikus.
INSTRUCTION
1.
Provide the patient immediately with a replacement Ikus.
HOTLINE
Notify Berlin Heart! 866.249.0128
14.9 Alarm circuit fault: buzzer remains off (or on)
•
Alarm circuit test failed - buzzer remains off!
•
Alarm circuit test failed - buzzer remains on!
WARNING
If this is the case, the Ikus will not generate an acoustic signal in an
alarm situation, or it will generate a wrong signal. Observe the
messages displayed in the message window carefully and look out for
the visual alarm signal in the display and operating panel. Do not
operate the Ikus without supervision!
A fault in the alarm circuit is discovered during the self-test for the alarm circuit or when
an alarm situation occurs. Depending on the type of fault, the message appears:
•
Alarm circuit test failed - buzzer remains off!
•
Alarm circuit test failed - buzzer remains on!
•
Acoustic alarm is not properly recognized
INSTRUCTION
1.
Provide the patient immediately with a replacement Ikus.
HOTLINE
Notify Berlin Heart! 866.249.0128
14.10 Backup computer faulty! Contact customer service.
Backup computer faulty! Contact customer service
.
NOTICE
Depending on when the fault occurs, the error message
Processor
down. Inform service!
may not appear due to unfavorable
resynchronization time response. In this case the drive will provide a
visual and acoustic signal, however, no error message will be
displayed.
One of the processors has failed. If the main computer fails, the message
Backup
computer started! Contact costumer service.
also appears.
INSTRUCTION
1.
Provide the patient immediately with a replacement Ikus.
1000721x09 Revision 8
178
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...