Chapter 14 Error Messages and corrective measures
14 Error Messages and corrective measures
This chapter describes all Ikus error messages and explains what measures should
be taken if an error does occur.
Whenever a message is displayed, always follow the exact instructions provided in
this instruction for use.
Keep calm!
It is necessary to be very observant as long as the cause of a message has not been
corrected.
Ikus treats all error messages with the same priority.
HOTLINE
Notify Berlin Heart! 866.249.0128
NOTICE
Some errors immediately re-trigger an alarm as long as they are still
active even after being acknowledged on the laptop. Also in this case
first mute the alarm on the handle in order to prevent permanent
retriggering of the acoustic alarm. Before acknowledging the alarm wait
for the error message and then take appropriate action.
When an error message occurs, the following happens:
•
An acoustic signal (2 different beep sounds) is emitted.
•
The indicator light on the control panel of the handle lights up.
•
A red border is shown around the field
Alarm off
in the monitor program display.
•
In the message window of the monitor program, a text is displayed informing
about the time of occurrence, the type of fault and the corrective measures which
must be taken. IMPORTANT: Always observe the instructions! In addition, some
of the more complex error messages contain an 8- or 16-digit binary code which
enables the service department to identify the exact cause of the fault.
What to do when an error message is shown
INSTRUCTION
1.
Check the status of the patient.
2.
Observe the filling and ejection behavior of the blood pump (visual check) over
several pump cycles!
3.
Carry out the appropriate measures and acknowledge the message.
Error message
Fig. 14-1
Example of an error message in the message window
1000721x09 Revision 8
171
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...