Chapter 13 Weaning and Explantation for BTR and BTT
Reduced rate (RR
x
)
Calculation
RR
1
Please enter:
RR
1
= WR 50 bpm + 0.75 x TWI ( ___ bpm) = ____bpm
RR
2
Please enter:
RR
2
= WR 50 bpm + 0.50 x TWI ( ___ bpm) = ____bpm
RR
3
Please enter:
RR
3
= WR 50 bpm + 0.25 x TWI ( ___ bpm) = ____bpm
Tab. 13-2
Reduced rate day 1 to day 3
10 ml / 15 ml pump Weaning Sequence
10 ml / 15 ml pump Weaning Sequence
Day 0
After confirmation of eligibility criteria, the following steps should be
performed under echo guidance
1
:
1. Administer unfractionated heparin (UFH) 75 units/kg x _____ kg =
_____ mg IV x 1 [max 5000 units].
2. After 5 minutes, reduce the pump rate step-wise from IR (_____ bpm)
to 30 bpm in increments of 5 bpm q5 min. After 5 minutes at 30 bpm,
reassess LV size and function.
3. After an additional 5 minutes (i.e. total time =
10 min
at 30 bpm), stop
the pump for
3 min
and reassess LV size and function. During pump
stoppage, use the manual pump to pump twice q30 seconds to
minimize the risk of thrombus formation due to blood stagnation while
Ikus is disconnected.
4. After 3-minute pump stop, reconnect pump to Ikus and resume pump
speed at IR(___ bpm).
Did the patient satisfy all 5 eligibility criteria
throughout the period of weaning and pump
stoppage?
ÿ
NO -STOP
ÿ
YES
Proceed
MD________
Tab. 13-3
10 ml / 15 ml pump weaning sequence
1000721x09 Revision 8
155
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...