Chapter 9 Implantation - surgical procedure
These sets could be neccessary in the following contexts:
• during implantation
• when replacing a blood pump
• when cutting off a piece of cannula (due to visible deposits or damaged cannula)
The cannula extension set / connecting sets thus guarantee that the blood pump and
cannulae can still be safely connected with one another and that the cannulae and
titanium connectors on the blood pump can still be visually inspected.
Each cannula extension set / connecting set consists of two cannula extensions /
connectors. Each cannula extension / connector comprises a double-sided titanium
connector to which a piece of tube is connected on one side.
Fig. 9-5
Cannula extension: titanium connector with tube section
9.3.2
Instructions for Use: Cannula extension set and connecting set
Cannula extension set
WARNING
If, on
further shortening
of the cannula, visual inspection of the titanium
connector on the blood pump is no longer possible: use the cannula
extension set.
Cannula extension set and connecting set
WARNING
The cannula extension set and the connecting set should only be used
if neccessary, since the basic risk of thrombogenesis and deposits
increases each time the cannula is extended.
Do not combine the connecting set with stage cannulae in such a way
that multiple diameters are bridged. Otherwise, the pump will not fill or
empty completely.
Secure each of the connections with at least 1 cable tie. Otherwise, the
connections may loosen over time and the cannula extension set / the
connecting set may become separated from the blood pump.
All effort should be made to minimize the manipulation and distortion of
the blood pumps and cannula during the removal of the cable tie(s) to
prevent mobilization of deposits.
If it is necessary to clamp any other part of the cannula (cannula
extension set / conecting set resp. ) that is not covered with velour,
cover the part that will be clamped with a gauze sponge
1000721x09 Revision 8
119
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...