Chapter 7 Commissioning the Ikus and setting parameters
INSTRUCTION
1.
Select
Drive pause
then press
<Enter>
to confirm. Confirm the decision in the
dialog window by pressing the
<X>
key or the
<1>
key. The Ikus will stop. The
view
Select operating mode
is displayed.
2.
Disconnect both tank units from the
Ikus
.
3.
Seal the driving tube connection sockets with seal plugs.
7.2.2 Selecting the operating mode (view Select operating mode)
Fig. 7-3
Select operating mode
INSTRUCTION
1.
In the operating room the
Ikus
should be reconnected to the mains immedi
ately.
2.
Select
Univentricular (UVAD)
or
Biventricular (BVAD)
with cursor, then con
firm the selected operating mode with
<Enter>
.
3.
In biventricular mode: the view
Pump size and single-step mode
appears (see
Fig. 7-7, page 100).
In univentricular mode, a connector seal test is first performed: the Ikus checks
whether the driving tube connector socket with the blue marking has been
sealed.
Univentricular operation: connector seal test
NOTICE
The Ikus will repeat the connector seal test up to 3 times. If the test is
still not successful, the system will switch itself off. Please contact the
emergency hotline.
HOTLINE
Notify Berlin Heart! 866.249.0128
INSTRUCTION
1.
The message
Please close right outlet
is shown. Check the driving tube con
nector identified by a blue marking. If the connector is open: use the seal plug
to close it.
2.
Move the cursor to
OK
and press
<Enter>
to confirm. The
Ikus
will test
whether the connector is properly sealed. If it is, the view
Pump size and sin
gle-step mode
is displayed. If not, the
Ikus
will repeat the connector seal test.
1000721x09 Revision 8
7.2
Intraoperative drive management
7.2.1 Disconnecting the tank unit from the Ikus
97
Summary of Contents for EXCOR Pediatric VAD
Page 2: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 4: ...This page was left blank intentionally 1000721x09 Revision 8 ...
Page 76: ...Chapter 5 Description Ikus This page was left blank intentionally 1000721x09 Revision 8 66 ...
Page 231: ...Chapter 16 Appendix 16 4 Sample copy Ikus Incoming Checklist 1000721x09 Revision 8 221 ...
Page 232: ...Chapter 16 Appendix 16 5 Implantation record form 1000721x09 Revision 8 222 ...
Page 233: ...Chapter 16 Appendix 1000721x09 Revision 8 223 ...
Page 234: ...Chapter 16 Appendix 1000721x09 Revision 8 224 ...
Page 235: ...Chapter 16 Appendix 1000721x09 Revision 8 225 ...
Page 236: ...Chapter 16 Appendix 1000721x09 Revision 8 226 ...
Page 239: ...Chapter 16 Appendix 1000721x09 Revision 8 229 ...
Page 240: ...Chapter 16 Appendix 16 7 Pump performance flow sheet 1000721x09 Revision 8 230 ...
Page 246: ...Chapter 17 EMC tables This page was left blank intentionally 1000721x09 Revision 8 236 ...