156
1015002x00
Identifying Errors and Troubleshooting
en
17.3 Driving Unit and Panel PC Alarms
17.3.1 Alarms
If an alarm condition develops, an alarm will be emitted:
17.3.2 Priority
Different alarms have differing priority levels:
high
,
medium
and
low
. Higher priority
alarms are signaled visually and audibly before lower priority ones. When alarms of
the same priority level occur at the same time, the most recent alarm will be shown on
the display first.
The user interface displays the higher-priority alarm on the left hand side. If multiple
alarms have the same priority level, the most recent one will be displayed on the left.
Responding to alarms by priority
Identifying alarm priority
Driving unit
Panel PC
External alarm
Message
Message and
instructions on
the display
Detailed mes-
sage and
instructions on
the user inter-
face
Forwarding of
high- and
medium-priority
alarms. Alarms
are emitted in
accordance with
the system used
at the hospital.
Visual alarm
Battery, system,
and mainte-
nance LEDs
Bell on the user
interface
Audible alarm
Audible alarm
(for exceptions,
see
section 17.3.8,
page 159)
—
Tab. 17-1
Alarms
high
Requires users to react immediately.
medium
Requires users to react in a timely fashion.
low
Requires acknowledgment by the user.
Priority
Driving unit
Panel PC
high
red flashing system LED
or battery LED
rapid tone sequence
every 2.5 seconds
red flashing bell, alarm
code on red background
Tab. 17-2
Identifying priority
Summary of Contents for EXCOR Adult
Page 2: ...2 1015002x00 en ...
Page 12: ...12 1015002x00 Table of Contents en ...
Page 14: ...14 1015002x00 Contact en ...
Page 66: ...66 1015002x00 Driving Unit en ...
Page 94: ...94 1015002x00 User Interface en Fig 7 7 Help window 1 Help display 2 Help button 1 2 ...
Page 122: ...122 1015002x00 Configuring and Operating the Driving Unit en ...
Page 126: ...126 1015002x00 Replacing the Driving System en ...
Page 128: ...128 1015002x00 Weaning and Explantation en ...
Page 134: ...134 1015002x00 Mobile Use of EXCOR Active en ...
Page 150: ...150 1015002x00 Baby Buggy Stroller Board en Fig 15 3 Guiding of the driving tubes ...
Page 154: ...154 1015002x00 Manual Pump en ...
Page 170: ...170 1015002x00 Identifying Errors and Troubleshooting en ...
Page 180: ...180 1015002x00 Maintenance Cleaning and Disinfection en ...
Page 194: ...194 1015002x00 Technical Specifications en ...
Page 198: ...198 1015002x00 Overview Parameters en ...
Page 206: ...206 1015002x00 EMC en ...
Page 208: ...208 1015002x00 Patient ID Labels en ...
Page 210: ...Abbrevations 210 1015002x00 en ...
Page 212: ...212 1015002x00 en ...