Application
Class I, functional test, direct current measuring method
Place the test sample onto an insulated surface.
Connection diagram
VDE 0701-0702, VDE 0544-4, VDE 0751-1
Circuit diagram
VDE 0701-0702, VDE 0544-4, VDE 0751-1
Insert the shock-proof plug of the portable test sample into the
test socket of the appliance tester.
The test sample is supplied with mains voltage and must be
tested in all switch positions (device functions). Please observe
inductive/capacitive circuits.
Class I, functional test, differential current measuring method
Measures the sum of all leakage currents
Connection diagram
VDE 0701-0702, VDE 0544-4, VDE 0751-1
Circuit diagram
VDE 0701-0702, VDE 0544-4, VDE 0751-1
Insert the shock-proof plug of the portable test sample into the
test socket of the appliance tester.
The test sample is supplied with mains voltage and must be
tested in all switch positions (device functions). Please observe
inductive/capacitive circuits.
Measurement
Make sure that the test sample is free of mains voltage.
Connect the mains plug of the test sample to the test socket of
the appliance tester.
Select a measuring
method.
Direct / differential current measuring method
Class I
Class II
Tap the “
Start
” button.
Tap the “
Start
” button.
As part of the functional test, the
IPE
measurement is carried out
automatically via the protective
conductor of the inserted shock-
proof plug of the test sample.
Use the test probe to the
following parts:
VDE 0701-0702/ VDE 0544-4
ICont
measurement
o
Accessible conductive parts
VDE 0751-1
ILeak
measurement
o
Active parts,
o
Accessible conductive parts
connected to the housing.
Tap the “
L<->N
” button.
Tap the “
L<->N
” button.
The
IPE
measurement is carried
out again via the protective
conductor of the inserted shock-
proof plug of the test sample.
Use the test probe to scan the
parts mentioned above again.
Tap the “
Stop
” button to stop the test.
The “
Test result
” screen opens.