Class II, ICont, alternative leakage current measurement
L and N conductors are bridged
Connection diagram
VDE 0701-0702, VDE 0544-4
Circuit diagram
VDE 0701-0702, VDE 0544-4
Insert the shock-proof plug of the portable test sample into the
test socket of the appliance tester.
The testing voltage is applied and the test sample must be tested
in all switch positions (device functions).
Use the test probe to scan all accessible conductive parts.
Measurement
Make sure that the test sample is free of voltage.
Connect the mains plug of the test sample to the test socket of
the appliance tester.
Select a measuring
method.
Direct / differential current
measuring method
Alternative leakage current
measurement
Tap the “
Start
” button.
Tap the “
Start
” button.
Use the test probe to scan all
accessible conductive parts of
the test sample.
Use the test probe to scan all
accessible conductive parts of
the test sample.
Tap the “
L<->N
” button.
Use the test probe again to
scan all accessible conductive
parts of the test sample.
Tap the “Stop” button to stop the test.
The “
Test result
” screen opens.