5.7 Measuring the polarisation index (PI) or the dielectric absorption ratio (DAR)
5225 / 04/2022 en
BENNING IT 100
31
5.7
Measuring the polarisation index (PI) or the
dielectric absorption ratio (DAR)
Requirements
• Please observe the requirements for measuring [
• Approved safety measuring lines
• Measuring ranges for insulating resistance measurement [
• Measuring point must be free of voltage
WARNING
Dangerous voltage
Danger to life or serious injury is possible due to contact with high electric voltage in case of
incorrect operation. During insulating resistance measurement, dangerous testing voltages
might be present at the device. These might also be applied to the measuring circuit if safety
measuring lines are contacted.
• Do not touch the bare measuring probe tips of the safety measuring lines or the bare
contacts of the optional alligator clips, Only touch the safety measuring lines in the area
intended for your hands.
Polarisation index (PI)
Polarisation index (PI) = R10‑min / R1‑min
• R10‑min: Measured insulating resistance after 10 minutes
• R1‑min: Measured insulating resistance after 1 minute
Dielectric absorption ratio (DAR)
Dielectric absorption ratio (DAR) = R1‑min / R30‑sec
• R1‑min: Measured insulating resistance after 1 minute
• R30‑sec: Measured insulating resistance after 30 seconds