5.1 Requirements for tests and measurements
Operation
5191 / 07/2021 en
BENNING CM 2-1
26
5
Operation
The device enables you to carry out various tests and measurements.
5.1
Requirements for tests and measurements
• Remove the device (safety measuring lines) from the measuring point before setting a
switch position on the rotary switch of the device.
• Only use approved safety measuring lines [
• Please consider sources of interference that might be present. Strong sources of
interference in the vicinity of the device might involve unstable readings and measuring
errors.
• Please observe the associated measuring ranges and measuring accuracies in chapter
page 22] for carrying out the tests and measurements.
DANGER
Maximum admissible voltage
Danger to life or serious injury is possible due to contact with high electric voltage.
• Use the device only in electric circuits of overvoltage category III with a conductor for a
maximum of 600 V or of overvoltage category IV with a conductor for a maximum of 300 V
to earth.