BENNING IT 115
Technical data
- 54 -
7.2 Low-impedance resistance / continuity test
7.2.1 Low-impedance resistance R LOW
Measuring range according to EN 61557-4: 0.16
1999
Measuring range R (
)
Resolution (
)
Accuracy
0.00
19.99
0.01
(3 % of the measured value + 3 digits)
20.0
199.9
0.1
(5 % of the measured value)
200
1999
1
Measuring range R+, R- (
)
Resolution (
)
Accuracy
0.0
199.9
0.1
(5 % of the measured value + 5 digits)
200
1999
1
Open-circuit voltage ............................... 6.5 V DC
9 V DC
Measuring current ................................... min. 200 mA for a load resistance of 2
Test cable compensation ........................ up to 5
Number of possible tests ........................ > 2000, with fully charged batteries / storage batteries
Automatic polarity reversal of the testing voltage.
7.2.2 Continuity test
Measuring range (
)
Resolution (
)
Accuracy
0.0
19.9
0.1
(5 % of the measured value + 3 digits)
20
1999
1
Open-circuit voltage ............................... 6.5 V DC
9 V DC
Short-circuit current ................................ max. 8.5 mA
Test cable compensation ........................ up to 5