Menu mode: Operation and setting
61
TM800_D00165_00_M_XXEN/06.2015
1. Exit
Back to the main menu
2. Edit table Use the arrow buttons to select the line "No" and confirm with the "
↵
" button.
Use the arrow buttons to set the external BMS bus address "Ext" and confirm
with the "
↵
" button.
Use the arrow buttons to set the internal BMS bus address "Int" and confirm
with the "
↵
" button. Use the arrow buttons to select insulation monitoring
device "Type" and confirm with the "
↵
" button.
TM800 always adds a new line at the end of the table. This line can be mod-
ified (e.g. 4 000 000 off). In this way other test addresses can be activated.
3. Clear table
Deactivate all addresses ("off").
Possible settings for "Type":
8.4.4.4
Settings menu 4: Measurement addresses
Currently has no function.
8.4.4.5
Settings menu 5: Digital inputs
Setting of the operational characteristics of the digital inputs IN1…IN.. (TM800 in combination with
BMI8/8 or BMI8/4 only. The following setting can be made individually for each input: "24V" (high) or
"0V" (low). When the input is set to "24V" an alarm message will be sent when the voltage at the input
is 24 V. When the input is set to "0V" an alarm message will be sent when the voltage is 0 V. It is for
this reason that you should always set unused digital inputs to "off".
1. Exit
Back to the main menu
2. Use the arrow buttons to select the operating mode for the digital input and confirm with the "
↵
"
button,select "0 V", "24 V" or "Off" using the arrow buttons. Press "
↵
" to ac-
cept the entry. Repeat the procedure to set more digital inputs.
3. Function Set the message text category to "neutral" or "medical", see chapter "Alarm texts for the
Syst. Meaning
107TD
Insulation monitoring device, e. g. 107TD47
IRDHx
Industrial insulation monitoring device, e. g. IRDH…
GFCI
Ground Fault Circuit Interrupter
LIM
Line Isolation Monitor
Off
Deletes the current line of the table
Alarm messages for medical gases are signalled by the red "ALARM" LED and the
buzzer sound. The buzzer can be set to mute. The buzzer sounds again every 12
minutes as a reminder.
Individual settings can only be made using the TMK-SET software.
1.Exit
2.Input: 001: 24V
Alarm at 24V
---Digital inputs---
1.Exit
2.Operat. principle
3.Function: neutral