System description
13
TM800_D00165_00_M_XXEN/06.2015
History memory with real-time clock to store 1000 warning and alarm messages
Variable operator and display modules with a freely programmable function
Easy integration with third-party systems, because operating theatre table controls, medical
gases, intercom systems are behind a closed foil surface
Control of third-party systems by flexible I/O modules with galvanic separation
Alarm LED at each input/output for fast diagnosis
Functions can be easily extended by adding I/O modules
Clearly defined structure thanks to an external and internal bus
Non-reflecting, multi-coloured foil surface
Optionally available with an antibacterial foil surface
3.3 Functionality of the TM800
3.3.1
LC display
The backlit features four lines of 20 characters. It assists medical and technical personnel during the
decision-making process with information that is always clear and unambiguous. Every alarm mes-
sage comprises three lines which appear spontaneously and three additional lines which can be dis-
played at the touch of a button. The fourth line contains status information (the number of
messages, test procedures, menu information).
Below the text display, three LEDs are arranged. They indicate: normal operation (green), warning
(yellow) or alarm messages (red).
Five buttons are available to acknowledge or to mute alarm and warning messages, for testing the
assigned devices and for the menu system.
Betrieb
Warnung
Störung