MK2000CBM

Installation Bulletin / Reference Guide

This document is intended as a reference guide to installing and setting an MK2000CBM series remote 

indicator. This document includes installation instructions and typical front plate display indications 

of the device. For complete details, including installation, setup, settings, and troubleshooting, refer to 

the LIM2010 user manual, document NAE2025010.

Only qualified maintenance personnel shall operate or service this equipment. These instructions 

should not be viewed as sufficient for those who are not otherwise qualified to operate or service 

this  equipment. This document is intended to provide accurate information only. No responsibility 

is assumed by BENDER for any consequences arising from use of this document. This device series is 

intended for use only with the BENDER LIM2010 Line Isolation Monitor.

Installation

Mounting

Two-gang plates provide four holes (Two at the top, two at the bottom) for screw mounting. 
Use the provided screws for mounting. Refer to figure 1 for dimensions.

Figure 1 - Two-gang wall plate dimensions in inches (mm)

Front Panel Display

1.  Green “SAFE” LED. Illuminates when the 

connected LIM2010 is in the normal 
condition.

2.  Yellow “HAZARD” LED. Illuminates 

when the connected LIM2010 is in the 
alarm condition.

3.  Yellow “OVERLOAD” LED. Illuminates 

when the connected LIM2010’s trans-
former overload alarm is active.

4.  TEST button. Initiates a self-test of the 

connected LIM2010.

5.  MUTE button with amber LED. Mutes 

the audible alarm when in the alarm 
condition. The amber LED indicates that 
the audible alarm has been muted.

6.  THC Digital Display. Shows the Total 

Hazard Current in real-time.

7.  Overload Digital Display. Shows the 

transformer overload percentage in re-
al-time.

1

4
5

2
3

6

7

Figure 3 - MK2000CBM diagram

Figure 2: Wiring Diagram, MK2000CBM

STW3 / STW4

CP-LIM2010 

Connector Plate

L1 from

transformer

secondary

L1 to

load center

To load center

RS-485 cable: twisted, shielded pair

MK2000CBM

120 

Termination

Resistor

SAFE

HAZARD

MUTE

ESC

TEST

RESET

MENU

LIM2010

To panel

ground bus

L1
L2

12VDC CM

A

B

RI1

K1/NC

K1/COM

K1/NO

SAFE

HAZARD

RI2

LIMGND

TEST

1S1

Z1/M+

1S2

Z2/M-

K2/COM

K2/NC

K2/NO

GND2

4.6“ (116.8)

4.55“ (1

15.6)

Wiring

MK2000CBM series remotes connect to a connector 
plate assembly utilizing a two-wire RS-485 connection. 
Ensure that the LIM2010 and corresponding connector 
plate have already been installed.

Refer to figure 2 below for wiring diagram. 

The 

MK2000CBM may only be used with type LIM2010 line 

isolation monitor.

! DANGER

• Disconnect all power before servicing.

• Reference NFPA 99 / CSA Z32 for

  Installation Standard.

HAZARD OF ELECTRIC SHOCK,

EXPLOSION, OR ARC FLASH

Applicable Devices

This document applies to MK2000CBM remote indicators only.