Benchmark BSH6000 Operation Manual Download Page 1



Operation Manual 

 

 
 

BSH6000 

Iso-Block  

Dry Bath Incubator 

 

 

 

Summary of Contents for BSH6000

Page 1: ...Operation Manual BSH6000 Iso Block Dry Bath Incubator...

Page 2: ...DATA ERROR BOOKMARK NOT DEFINED CHAPTER 3 OVERVIEW 3 1 STRUCTURE OVERVIEW 3 2 KEYPADS 4 3 DISPLAY 4 CHAPTER 4 OPERATION 5 1 TEMPERATURE AND TIME SETTING 5 2 ADVANCED TIMER SETTING ERROR BOOKMARK NOT D...

Page 3: ...lification of DNA the pre denaturation of the electrophoresis blood serum coagulation and a variety of laboratory procedures Features Large digital display of time and temperature Extremely accurate t...

Page 4: ...humidity 70 Power supply AC115V or 230V 50 60Hz 2 Technical Data Item No BSH6000 Temperature range Ambient Room Temp 5 C 105 C Timer Max 99hr 59min Temperature accuracy 0 1 C Temperature uniformity 0...

Page 5: ...er focuses on the introductions of the structure keypads and key functions of the instrument as well as preparatory work prior to initial operation 1 Structure overview 2 Power supply Fuse On Off butt...

Page 6: ...m Temp Time h m Block A Block A Block B Block B Block A Mode Block A Block B Time Start Stop Temp Time Mode Temp Time Time Start Stop TM 3 Display Mode Start Start Stop key After the temperature and t...

Page 7: ...y are actual settings real time temperature and time remaining The example shows an actual temperature of 28 5 C and 35 minutes remaining c To adjust the set temperature press or keys until the desire...

Page 8: ...even once time has expired and oVEr is showing on the time display This setting can be adjusted so that the heater will shut down once time has expired To adjust the advanced timer setting pres the M...

Page 9: ...A a Power on the machine and ensure that the temperature on the display is less than 35 C b Inject olefin oil into one of the cone shaped wells and then put a thermometer thermocouple into this well m...

Page 10: ...nfirm the input value d Then the instruments will the automatically begin to heat to 100 Once 100C is reched and held as constact the decimal begins to flicker After 20 minutes the user can read the a...

Page 11: ...se On Off button broken Change button Others Contact the supplier 2 The actual and displayed temperatures are quite different Broken sensor or loose contact of the block Contact the supplier 3 OPEn di...

Page 12: ...ual Annex 10 Annex Wiring Diagram for reference only KEY a nd Display PT1 Power HEAT1 HEAT2 PT2 MAIN BOARD FAN2 FAN1 Key andDisplay L PE N HEAT PT100 B lock1 HEAT PT100 B lock2 Fan Fan Switch PORT1 J3...

Page 13: ...Memo...