Belmont SP-CLEO II Operating Instructions Manual Download Page 6

- 2 -

1. SAFETY PRECAUTIONS

WARNING

CAUTION

9. Be sure to mount the mirror cover.

The mirror cover for the dental light must be attached whenever the light is used. Touching the light 

bulb may result in burns.

Refer to the operating instructions on the dental light for details.

10. Be sure to turn off the power when replacing the light bulb.

The power must be turned off when replacing the dental light bulb. Failure to heed this warning may 

result in electrical shock.

The specified halogen light bulb must be used.

The bulb and the bulb holder may be hot. Wait until they become cool before replacing the bulb.

Do not touch the halogen bulb with bare hands.

Refer to the operating instructions on the dental light for details.

11. Do not place articles weighing 3 kg or more on the instrument table.

Do not place articles weighing 3 kg or more on the instrument table. Failure to heed this warning may 

result in breakage, malfunction of the table, or an accident.

12. Do not apply excessive force to the arms.

Do not mount or apply unnecessary force to the unit arms or the armrests of the patient chair. Failure 

to heed this warning may cause an accident or the unit to tip over.

Persons other than dentists or other dental professionals must not use the system.

2. Check safety before operating the product.

Before operating the system, be sure to check that all parts operate normally and safely, and that there are no 

obstacles in the area.

1. The system must be operated or handled by a qualified dentist or by dental staff under the supervision of a 

3. Pay attention to the patient and children.

Be careful to avoid accidents due to mischief or careless operations by the patient (children in particular). 

Keep an eye on the patient.

4. Stop operation of the system when feeling something is wrong.

During inspections, pay attention to any looseness, wobbling, inclination, vibration, abnormal sounds, 

temperatures, or odors. If something is out of place, even slightly, immediately stop operating the system.

5. Do not strongly hit or scrape the system.

Do not strongly hit or scrape the system. Failure to heed this caution may result in breakage of the cover or 

malfunction of the system.

6. Be sure to operate the switches by hand.

Except for the footswitch, be sure to operate the switches by hand. Operation of switches by means other than 

hands may result in breakage or system malfunction.

7. Take caution when using water other than city water.

The system is designed for use with city water. Use of water other than city water (such as water-line 

sterilization system) may result in system failure.

Summary of Contents for SP-CLEO II

Page 1: ...Thoroughly read this Instruction Manual before use and use the SP CLEO II according to the instructions Keep this Instruction Manual carefully and read it again whenever it is needed This product is a...

Page 2: ...VERALL VIEW AND MAJOR COMPONENTS 4 DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS 5 CARE AND MAINTANANCE 6 STORAGE LIFETIME DISPODALAND RESTRICTIONS OF USE 7 BEFORE ASKING FOR REPAIRS 8 DIMENSIO...

Page 3: ...wed to operate or handle the product unless he she is so instructed The product is supplied together with the handpieces like electric micromotor air turbine and or motor scaler and so on Environmenta...

Page 4: ...Fiber optic handpiece light on off Rotation mode select Micro motor Forward Reverse select Syringe Function Store Minus Plus Bowl flush Cupfiller Dental light on off Dental light mode selection Servic...

Page 5: ...cidental contact with the equipment 3 Do not attempt to disassemble repair or modify the equipment Only authorized personnel should disassemble repair or modify the equipment Failure to heed this warn...

Page 6: ...dentists or other dental professionals must not use the system 2 Check safety before operating the product Before operating the system be sure to check that all parts operate normally and safely and...

Page 7: ...r Without the cover there is a possibility of injury caused by the scaler tip 14 Observe the following when handling handpieces Change the rotation speeds of the micro motor and detach the burr of the...

Page 8: ...rupture the clean water system 24 Do not use water for the clean water system optional other than purified water distilled water or pure water 4 Function of the flow adjustment valve knob The functio...

Page 9: ...not put hands or feet into the gap between the backrest and the seat Take care not to be trapped by the armrest Do not operate the chair with the right armrest swung out 180 degrees Take care not to b...

Page 10: ...ater heater Bowl flush About 5 minutes for cleaning water in cupfiller line Flushout procedure Handpiece line Pick up one handpiece out of the holder and hold it over the cuspidor bowl with the turbin...

Page 11: ...Instruments Holder Cuspidor Bowl Doctor Control Panel Doctor Table Balance Arm Doctor Instruments Holder Manual Switch Auto Switch Foot Controller Base Mount Type 3 OVERALL VIEW AND MAJOR COMPONENTS...

Page 12: ...Optional Dental Light Sensor Manual Switch applicable only for AL 520 AL 820S and 900 Dental light type 920 Docotor Table Section Cuspidor Section Holder Type Place Holder Type Cart Type Air Master V...

Page 13: ...he armrest to the former position observe the procedure which is opposite to that of lock release That is turn the armrest to the normal working position and then lock the armrest in position by pushi...

Page 14: ...seconds because the memorized position in auto mode may be changed 1 4 Stick Switch Manual Switch 1 Seat Lifting Keep depressing the manual mode stick switch to upward until the seat is lifted up to...

Page 15: ...lamp seizure and for power saving 1 120 seconds after steady state display appeared since the main switch power was turned on 2 30 seconds after normal operation The indicator is restored when any ope...

Page 16: ...he handpiece is actuated by picking it up from the handpiece holder and operating the foot controller See pages 25 to 30 for operation of foot controller micromotor and electric scaler Only one handpi...

Page 17: ...adjustment while the brake release button is pressed CAUTION Doctor table front arrow view Doctor table front arrow view With each one of these knobs the flow rate decreases when the knob is turned c...

Page 18: ...een brightness decreases when the brightness control dial is turned downward NOTICE Turn off the light when the filmviewer is not used Water jets out when the button with W mark is pressed Air jets ou...

Page 19: ...on is depressed Pedal The pedal depressing extent can control the turbine rotation speed and air scaler output The motor rotation is activated when the pedal is depressed The rotation increases when t...

Page 20: ...nsing position or treatment position each time when this switch is pressed Cuspidor flush switch Switch for flushing cuspidor bowl Cupfiller switch Switch for filling water to the cup Dental light swi...

Page 21: ...hair Press to recline the backrest Press to get up the backrest Chair preset switch Chair auto return switch Set the chair height and backrest angle in desired position s and actuate the chair by eith...

Page 22: ...e this switch cannot be used depending on the unit specification and the type of dental light Switch for on pff the handpiece coolant spray When a handpiece is picked up and this switch is pressed eit...

Page 23: ...itch for switching the micromotor rotation direction Pressing this switch each time changes the direction between forward clockwise rotation and reverse counter clockwise rotation Forward rotation is...

Page 24: ...4 DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS Use this switch for storing data and so forth 1 Store micromotor preset mode See page 45 for the setup procedure 2 Store preset timer 3 Store mi...

Page 25: ...mit changes in three steps or 5 steps Upper limit of rotation in case of three steps 10000 20000 40000 rpm Upper limit of rotation in case of five steps 5000 10000 20000 30000 40000 rpm See page 28 fo...

Page 26: ...e selected mode is displayed on the indicator either PERIO ENDO or SCALING For switching the mode use the switch or switch on the membrane switch panel The EMS scaler does not provide this mode select...

Page 27: ...tup Contents and sequence of function switch setup are described below Setup and operation of dental timer Selection of doctor number Selection and operation of flushout Selection and change of key to...

Page 28: ...ch is pressed 4 When the timer time is reduced to 00 00 the alarm sounds to inform that the set length of time has elapsed Procedure for setup of preset timer 1 Press the function switch once 2 Set th...

Page 29: ...ou want to change setup of the light pack residual lighting time please contact your dealer or our office Procedure for setup 1 Press the function switch twice 2 Press one switch from and for selectio...

Page 30: ...the tone To increase the frequency To decrease the frequency To change the frequency to initial setup Precautions for change to frequency Ensure to press the switch on finishing the frequency change...

Page 31: ...six times 2 Press a desired chair setup switch from and 4 DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS Selection of number of steps of micromotor limit rotation The number of steps of micromo...

Page 32: ...auses the micromotor to run at a preset rotation regardless of the foot controller pedal depressing extent The micromotor preset mode permits memory setup of SET1 to SET3 Procedure for setup 1 Pick up...

Page 33: ...turned clockwise and the flow rate increases when the knob is turned counterclockwise Water heater switch Switch for turning on off the cupfiller water heater The cupfiller water is heated to warm wa...

Page 34: ...e and the flow rate increases when the knob is turned counterclockwise 3 3 Vacuum handpiece Suction begins when the vacuum handpiece is taken out of the holder In case of the central vacuum system suc...

Page 35: ...a is wetted wipe off the moisture as soon as possible It will rust otherwise Wipe the resin cover with a wet and soft cloth Chair Sterilization Use FD333 made by Durr or ethanol as sprayed to a soft c...

Page 36: ...r is provided with sharp edges for easily catching cotton or the like Therefore be careful at the occasions of cleaning Oil mist separator optional The oil mist separator optional collects the oil con...

Page 37: ...irection A The nozzle can be removed from the syringe Remove dirt with tap water before sterilization Apply equal thin coat of vaseline to two O rings after sterilization To return the nozzle to the o...

Page 38: ...r cleaning Use of strong acid detergent or alkali pipe detergent may corrode metals Never use detergent of such a nature Make sure to turn off the main switch before dismounting or mounting of the cus...

Page 39: ...ody section Disassembling and cleaning procedure 1 Pull the cap section and body of the vacuum handpiece to outside and carry out disassembly as shown in the figure Also remove the valve from the slid...

Page 40: ...moved when the joint is unlocked by turning it counterclockwise Therefore they can be cleaned with running water For cleaning the vacuum hose and saliva ejector hose turn off the main switch for the u...

Page 41: ...rning it clockwise Drain the water accumulated in the air compressor with the drain valve opened once a week 5 CARE AND MAINTANANCE 4 Junction box section Cleaning of product exterior Apply dry wiping...

Page 42: ...e to the closed position by turning it clockwise if it is anticipated that the unit will not be used for a long period of time for recess or the like 4 Air main valve Turn the air main valve to the cl...

Page 43: ...ol valve or knob for handpiece or the like is closed Open the valve or knob Vacuum suction does not work Vacuum pump power is not on Turn on the power Filter of solid collector is filthy Clean the fil...

Page 44: ...520 AL 820S 900 Dental Light Type 920 Working environment Temperature 10 to 40 Humidity 30 to 75 Atmospheric pressure 700 to 1060hPa Maximum Load 135kg Classification Operation mode Short time operati...

Page 45: ...594 460 398 1710 Base mount type Unit mm Dimensional tolerance 10 41 400 570 1871 848 530 296 959 822 188 584 750 1000 483 748 594 1710 Max 2456 636 1939 for 920 light Cart type 8 DIMENSION AND SPECI...

Page 46: ...ling shall be executed Malfunction may arise if the cup is of another material grade such as stainless steel and plastics or if the paper cup is of dark color or pattern Cupfilling may not be executed...

Page 47: ...ice if the main switch cannot be turned off or if the main valve cannot be closed Product s moving parts No abnormal noise or the like shall be produced from product s moving parts when the product is...

Page 48: ...2 1 1 Higashishinsaibashi Chuo ku Osaka 542 0083 Japan TEL 81 6 6213 5945 FAX 81 6 6212 3680 TAKARA BELMONT CORPORATION Printed in Japan 2011 09 NOTE BOOK NO 1E00ZXB0...