- 28 -
No.1 SET1
10000rpm
Preset revolution (SET1)
Revolution (rpm) of
preset revolution (SET1)
For changing
the revolution
For storing
min.
sec.
=LINK F=E
(-)=IND (+)=LINK
(Example) Case of interlock
Non-interlock (individual) Interlock
Selection of interlocked action of cupfiller and bowl flush
Selection is permitted from two procedures.
Case of non-interlock
Flushout of the cuspidor bowl is not implemented even after
start-up of cupfiller.
Case of interlock
Flushout of the cuspidor bowl is implemented simultaneously
with start-up of cupfiller.
[Procedure for setup]
1. Press the function switch nine times.
2. Press the desired setup switch.
Non-interlock --- switch
Interlock --------- switch
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
2-18. Setup of micromotor preset mode
The micromotor preset mode causes the micromotor to run at a
preset rotation regardless of the foot controller pedal depressing
extent.
The micromotor preset mode permits memory setup of SET1 to
SET3.
[Procedure for setup]
1. Pick up a micromotor from the holder.
2. For memory setup of preset mode SET1, press rotation mode
select switch until SET1 is displayed on the indicator.
3. SET1 is displayed on the indicator, and at the same time, the
rotation set in SET1 is displayed.
The micromotor runts at this displayed rotation (value for
reference).
4. To change the rotation, press the switch or switch.
The numerical value displayed on the indicator changes.
5. To store the rotation after change, press the switch.
The set value before change is restored if the main switch is off
without pressing this switch.
6. Memory setup of each of preset mode SET2 and SET2 can be
implemented by observing the procedure described in step 2 and
subsequent above.
Do not change the rotation while the micromotor is running.
CAUTION
9