Balance Arm Brake
Release Button
2-8. Balance arm brake
For base mount type
Use this button for releasing the balance arm brake to raise or
lower the table.
Keep this button pressed while raising or lowering the table.
For cart type
Loosen the column knob to raise or lower the table.
Fix the table in place by firmly tightening the column knob
after adjustment.
Do not place a load of over 3 kg on the table, or table damage,
malfunction and failure may be caused.
NOTICE:
Noise might be produced due to impact at the time of table stop if the table is moved
upward or downward without pressing the brake release button. Make sure to carry out table height
adjustment while the brake release button is pressed.
CAUTION
E A
E A
①
②
Doctor table front arrow view
Doctor table front arrow view
* With each one of these knobs, the flow rate decreases when
the knob is turned clockwise, and the flow rate increases
when the knob is turned counterclockwise.
* The flow rate control knobs are provided for increasing/
decreasing the flow rate, and they are not stop valves.
Note that the knob will idle if it is shut off with force.
Increase
Decrease
4
6
7
8 9
2-6. Handpiece spray flow rate control
2-7. Syringe spray water and air control
The syringe water and air flow rate can be adjusted by turning
knobs of the doctor section.
Blue cap ------ Water
Yellow cap --- Air
The handpiece spray flow rate can be adjusted by turning
knobs (with blue caps) of the doctor section.
Column Knob
- 13 -
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS