Handpiece Pressure Gauge
Main Switch
ON OFF
MAIN
Lit in green
Indicator
4. DESCRIPTION OF OPERATION AND FUNCTIONS OF COMPONENTS
2. Doctor Unit Section
2-1. Main switch
E A
When the main switch is turned on, the MAIN LED on the doctor
membrane switch panel lights up in green. This state permits
motion of the unit, chair and dental light.. The unit can be moved
even while the chair is OFF.
When the main switch is turned on, all the lamps on the doctor
membrane switch panel light up, and the following is displayed on
the indicator.
A few seconds later,
is displayed and all the functions can be operated.
* The lamps turn off automatically if the equipment is not used for a
certain length of time. But they light up again when any operation
is performed.
E A
Indicator off function
The indicator is off if no operation is performed for a fixed length of time. This function is provided
for preventing fluorescent lamp seizure and for power saving.
1. 120 seconds after steady state display appeared since the main switch (power) was turned on.
2. 30 seconds after normal operation
The indicator is restored when any operation is attempted.
2-2. Indicator
Indicator abnormality and troubleshooting
There are cases where the display on the indicator becomes abnormal due to noise.
In such a case, turn off the main switch and turn it on again about 5 seconds later.
* However, there is a possibility where reset does not occur if noise is continuing.
Eliminate the cause for the noise, and then turn on the power again.
If it is not possible to identify the cause for the noise, please contact your dealer or our office.
E A
If any error arises to the memory, the buzzer sounds and [EEPROM READ ERR] is displayed on the
indicator. Once turn off the main switch and then turn on the main switch again. If an error message is
still displayed, please contact your dealer or our office.
* This doctor unit section is actuated by compressed air. Therefore, ensure that the air compressor is
working.