29
Management of maintenance and inspection of medical equipment
should be implemented by the user (medical institution).
For safe use of the product, it is necessary that inspection should be
conducted in the specified frequency on the items described below.
If the chair does not work normally , immediately stop using the chair,
turn off the power, unplug the power plug, and contact your local
authorized Belmont dealer.
Guide for daily maintenance and inspection
(Maintenance and inspection by user)
Inspection method
and diagnosis
Influence if inspection
not conducted
Maintenance required in
case of nonconformity
Confirma-
tion of the
cancel
function for
auto
operation
Power
supply
cable
Function
of each
switch
Moving
parts
Oil leakage
Stain
removal of
external
parts
Before
start
Before
start
Before
start
Before
start
Before
start
After
closing
Unexpected personal
injury and troubles may
arise due to motion of
the chair during medical
treatment and due to
pinching between doctor
section and chair.
Accident or defect of
the equipment.
Defect of movement
may cause an accident.
Accidend or defect of
the equipment.
Defect of the equipment.
Discoloration, transfor-
mation or breakage of
resin parts may arise.
Contact your local authorized
Belmont dealer if any abnor-
mality arises.
Clean and rearrange the
cable. For replacement,
contact your local authorized
Belmont dealer.
Contact your local authorized
Belmont dealer if any abnor-
mality arises.
Contact your local authorized
Belmont dealer if any abnor-
mality arises.
Contact your local authorized
Belmont dealer if any abnor-
mality arises.
Clean up according to the
“Cleaning Procedures“ in this
instructions for use.
Operation should stop if any of
the stick switches is pushed
during auto operation.
Power supply cable should not
be abnormally bent, pintched or
damaged, and the plug should
be firmly inserted to the power
socket and not covered with
dust.
Each switch for chair operation
should work correctly.
Make sure that the chair does
not make any abnormal noise
during automatic and manual
movement.
Make sure that there is no
hydraulic oil on the floor or chair
base leaking from the cylinder.
Visual check the attachment of
dust, chemicals or drug and
grasp the total product condi-
tion.
Frequency
No.
Item
WARNING
Always refer to this Guide before conducting daily maintenance and
inspection.
If you do not conduct daily maintenance or inspection, use of the
product may result in injury or damage to nearby devices.
1
2
3
4
5
6
Maintenance and Inspection