Recommended separation distances between
distances between
between
Portable and mobile RF communications equipment and the PHOT-X
Ⅱ
s 505
x-ray
The PHOT-X
Ⅱ
505
x-ray is intended for use in an electromagnetic environment in which radiated RF disturbances
for use in an electromagnetic environment in which radiated RF disturbances
are controlled. The customer or the user of the PHOT-X
Ⅱ
s
505
x-ray can help prevent electromagnetic interference
by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and
the PHOT-X
Ⅱ
s
505
x-ray as recommended below, according to the maximum output power of the communications
equipment.
Rated maximum output
power of transmitter
W
Separation distance according to frequency of transmitter
m
150 kHz to 80 MHz
d
= 1.2
√
P
80 MHz to 800 MHz
d
= 1.2
√
P
800 MHz to 2.5 GHz
d
= 2.3
√
P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance
d
in metres
metres
(m) can be estimated using the equation applicable to the frequency of the transmitter, where
P
is the maximum output
power rating of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800MHz, the separation distance for the higher frequency range applies.
NOTE 2 These quidelines may not apply in all situations. Electromagnetic propagation is affected by adsorption and
re
fl
ection from structures, objects and people.
-