Belmont IO 5000TA Installation And Operation Instructions Manual Download Page 17

[11] MAINTENANCE AND INSPECTION

11-1.Guide for daily maintenance and inspection (Maintenance and inspection by user)

Management of maintenance and inspection of medical equipment should be implemented by the user (medical 
inspection). In case the user does not implement such management, it is permitted that such management is 

outsourced to a qualified entity such as a medical equipment repair company.
For safe use of this product, it is necessary that inspection should be conducted in the specified frequency on 

the items described below.

No. Item

Frequency

Inspection Method and diagnosis

Influence if inspection is
not conducted

Maintenance required
in case of
nonconformity

1

Appearance of Lamp
Cover

Before start

Check deformation, scratches by
eyes

Deterioration of optical
performance

Change a lamp cover

2

Lamp cover is firmly
attached or not

Before start

Pull the lamp cover. Make sure that
the lamp cover doesn't fall

Lamp cover may fall.

Reattach a lamp
cover

3

Light bulb

Before start

Check whether the light brinks or not
due to the poor connection.

Light bulb turns very hot.
This could shorten bulb
life.

Securely insert a light
bulb into the bulb
holder

4

Light switch

Before start

Set Sensor/Manual switch to Manual.
Confirms the light turns on.

Light doesn't function.

Replace a light bulb

6

Intensity switch

Before start

Change intensity.

It may cause the difficulty
at dental practice.

7

Functionality of
Balance Arm

Before start

Swing the balance arm up/down.
Swing the balance arm right/left.
Make sure the balance arm stops and
holds its position.

Light head doesn't stay at
the desired position.

Adjust the tension of
the balance arm.

8

Vertical rotation of the
Light head

Before start

Rotate the light head up/down. Make
sure the light head stops and holds its
position.

Light head doesn't stay at
the desired position.

Adjust the tension of
vertical rotation.

9

Light head angle

Before start

Confirms the light head is vertically
aligned.

Light head doesn't stay at
the desired position.

Adjust the angle of the
light head.

10

Light head rotation

Before start

Rotate the light head right/left. Make
sure the light rotate 160° to each
direction.

Light head rotates freely.
This may cause the
snapping of a wire.

13

Care exterior

Before start

Chemical and dirt on the product
exterior must be cleaned.

Discoloration and
deterioration to the
exterior, and corrosion and
rusting to metallic
components may arise.

Executes wiping in
accordance with
"Cleaning
Instructions"

Other 1

Once every
week

Make sure that no abnormal noise
occurs when the product is operated.

Light may not function
right.

Other 2

If the light has not been used for a
long time, make sure the light
functions correctly and safely.

If the light
malfunctions, Contact 
your dealer or our office

As needed

Touchless sensor

Before start

Set Mode Selection switch to Sensor.
Confirms the light turns on.

Light doesn't function.

Clean the sensor
surface.

5

11

12

Turn off the light.Contact 
your dealer or our office

Contact your dealer or 
our office

Turn off the light.Contact 
your dealer or our office

-11-

Summary of Contents for IO 5000TA

Page 1: ...NSTALLATION AND OPERATION INSTRUCTIONS DENTAL LIGHT IMPORTANT After installation is completed check all the bolts screws and fasteners to confirm that they are securely fastened TYPE AL 820S IO 5000TA...

Page 2: ......

Page 3: ...e 4 8 ADJUSTMENT OF THE BALANCE ARM 8 1 Adjustment of the tension of the balance arm 5 8 2 Adjustment of the angle of the light head 5 8 3 Adjusting the tension of the light head 6 8 4 Balance arm adj...

Page 4: ...peration and accidents In case of disposal of equipment When disposing the dental light appropriately dispose complying with all current applicable regulations and local codes In EU area EU directive...

Page 5: ...ould be conducted by an authorized installation service personnel only WARNING The followings are prohibited To modify this equipment To use the equipment under any failure condition To use the equipm...

Page 6: ...ctly connected During use Do not use the light longer than required for examination or treatment Always watch the patient and the equipment to make sure nothing is wrong If anything wrong is observed...

Page 7: ...ht Pole Joint Balance Arm Joint Arm Joint Cover Balance Arm Arm Joint Cover Yoke Joint First Yoke LED Second Yoke Intensity Switch Handle Left Touchless Sensor Switch Reflector Cover Light Bulb Cover...

Page 8: ...ernate Intensity Selection Momentary SV2P Power Input 1 2 CN1 CN8 VH4P 1 4 3 2 1 2 Red Black Brown Setting of PCB Dip Switch SW1 1 2 3 4 OFF OFF ON ON OFF OFF OFF 4 3 2 1 SW2 Off Disable On Enable Def...

Page 9: ...n instruction of the unit Light Cable Light Head Light Pole Button Screw M6x8 Light Pole Joint 6 3 Connect the wire harness Fig 6 3 Attach the connectors of the light cable onto the dental light PCB a...

Page 10: ...ce and hold your hand within the reaction zone for 2 seconds The light intensity will be automatically reduced to the composite mode To switch back to the normal treatment mode place your hand by the...

Page 11: ...g the socket screw will damage the arm 2 Position the angle of the balance arm so the head angle adjustment nut becomes visible just under the slot C Turn the nut and adjust the tension with the adjus...

Page 12: ...w to adjust 1 Loosen the four M5 socket head screws E with the supplied 2 5mm Allen key When adjusting make sure that the socket head screws are loosened Failure to do so may cause damage to the threa...

Page 13: ...end of the upper side of the light pole joint and move the balance arm in place 7 Loosely tighten both top and bottom M8x20 D screws 8 Tighten both top and bottom M8x20 D screws and adjust the lateral...

Page 14: ...y after the lamp goes off Wait until they get cool down Do not touch glass with bare hand Halogen bulb surface must be clean Oil or body moisture will affect bulb span of life If glass surface is touc...

Page 15: ...the reflector cover 1 Insert the cut out of the reflector cover into the gap at the handle right See Fig 9 4 1 2 Be careful not to pinch the wires when attaching cover See Fig 9 4 2 3 Hold the reflect...

Page 16: ...d be taken to prevent scratching reflector surfaces as this will degrade the performance of the light CAUTION All surfaces can be cleaned with DURR FD333 cleaner Spray the cleaner DURR FD333 on cloth...

Page 17: ...ore start Swing the balance arm up down Swing the balance arm right left Make sure the balance arm stops and holds its position Light head doesn t stay at the desired position Adjust the tension of th...

Page 18: ...aken upon checkup stated above then stop using the unit turn off the main switch and contact your dealer or our of ce Unit main switch is not on Turn on the unit main switch The light does not light u...

Page 19: ...ser of the IO5000TA should assure that it is used in such an environment Immunity test IEC 60601 test level Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000...

Page 20: ...tromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 8...

Page 21: ...istance according to frequency of transmitter m 150 kHz to 80 MHz d 1 2 P 80 MHz to 800 MHz d 1 2 P 800 MHz to 2 5 GHz d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 1...

Page 22: ......

Page 23: ......

Page 24: ...4RC0 2 1 1 Higashishinsaibashi Chuo ku Osaka 542 0083 Japan TEL 81 6 6213 5945 FAX 81 6 6212 3680 TAKARA BELMONT CORPORATION Takara Belmont U K Ltd Belmont House One St Andrews Way Bow London E3 3PA U...

Reviews: