* In the Junction Box
Water Filter
Air Filter
Drain Valve
1
Junction Box Section
3
CARE AND MAINTENANCE
Cleaning air filter drain valve and discharging water from air compressor
• Drain valve is used to discharge water from the air
filter.
• Turn the drain valve counterclockwise to discharge
water collected in the air filter once a week at least.
• If water enters the unit, the air turbine, air motor or
syringe, etc., may become defective. Be sure to
turn the drain valve clockwise to close the valve
after discharging water from the air filter.
• Open the drain valve of the air compressor to
discharge collected water once a week.
Filter Replacement
• The water filter in the junction box needs to be
replaced at least once a year.
• The air filter in the junction box needs to be
replaced at least once every three years.
Contact your local service representative for
replacement.
* Water filter : Sintered metal element
: Rated filtration 100
"
m
Air filter : Sintered mertal element
: Rated filtration 40
"
m
1
Product Exterior
4
Cleaning and disinfection of product exterior
• Clean the metallic parts with a dry soft cloth.
Wipe off water immediately if water is put on the product. Water may cause rusting.
• Clean the resin parts with a wet soft cloth.
• Use FD333 or FD366 made by Durr or ethanol for cleaning and disinfection of the product exterior.
•
Wipe off water and residual disinfectant immediately. This could cause corrosion, damage or
incorrect operation of the unit.
•
Immediately wipe off any water spills or leakage on the floor. This could cause damage to the
product, decreased strength of the floor may lead to physical injury including fall, or property
damage.
CAUTION