Cuspidor Unit Section
2
CARE AND MAINTENANCE
Cleaning must be done within 1 hour after use. Replace with a new vacuum handpiece and saliva ejector
handpiece for following cases.
• Any waste material can not be removed by clogged hole.
• Contamination and solid material attached to vacuum handpiece and saliva ejector can not be removed.
CAUTION
CAUTION
29
Sterilization VH-18Vacuum Handpiece
【
Assembly before sterilization
】
Vacuum Tip/Vacuum Handpiece/Saliva Ejector
Handpiece can be autoclave. Vacuum handpiece and
saliva ejector handpiece have to assemble before
autoclave.
When assembling the vacuum handpiece, align the
two surfaces; a flat surface of the cap and the surface
of the sliding groove of the body. Then, slide the cap
into the body straight.
* Do not insert it twisted.
Caution when assembling
vacuum handpiece before
sterilization
Align the
two surfaces
After assembly
Cap
Body
Insert the cap
straight
down
Complete insertion
with two surfaces
tightly aligned
Confirm both surfaces aligned
After sterilizing the vacuum tip, keep it in the sterilization
pouch and store in a dark and cold place.