Belmont CREDIA G1 Operating Instructions Manual Download Page 31

(Holder Type)

 

(Rod Type)

Continental Type

1  

Doctor Unit Section

1

CARE AND MAINTENANCE

 Doctor Table Control Panel (Membrane Switch)

 Handpiece Hose

• Use FD333 or FD366 made by Durr or ethanol as

sprayed to a soft cloth for cleaning and disinfection.
Wipe off with dry soft cloth to dried up of the doctor
table control panel after cleaning and disinfection.

When the surface of the operation panel is cleaned 

with disinfectant, etc., wipe off disinfectant 

completely. If it penetrates into the back of the 

sheet, the membrane switches may malfunction.

• Use FD333 or FD366 made by Durr or ethanol as

sprayed to a soft cloth for cleaning and disinfection
of the handpiece hose.

Observe the instructions given in the package 

insert and Instruction Manual included with the 

handpiece to clean it.

NOTICE

CAUTION

26

Summary of Contents for CREDIA G1

Page 1: ...keep this instruction manual near the product for future maintenance Refer this manual as needed If you have any questions about this Manual or this product please contact us If manual becomes unread...

Page 2: ......

Page 3: ...RTS 22 23 1 DOCTOR UNIT SECTION 22 2 CUSPIDOR UNIT SECTION 22 3 JUNCTION BOX SECTION 23 CARE AND MAINTENANCE 24 31 1 DOCTOR UNIT SECTION 24 2 CUSPIDOR UNIT SECTION 27 3 JUNCTION BOX SECTION 31 4 PRODU...

Page 4: ...n be used with this Product Use the handpieces described on page 36 Chair that can be used with this Product Use the chair described on page 36 Dental Light that can be used with this Product Use the...

Page 5: ...outlet water Service outlet air Service outlet water flow control Water heater Chair preset 2 To recline the backrest To raise the chair Chair manual control Chair last position Chair auto return Cha...

Page 6: ...me 2 Be sure to turn off the main switch upon completion of work or during work breaks Be sure to turn off the main switch upon completion of work or during work breaks This prevents incorrect operati...

Page 7: ...o unexpected accidents 4 Discontinue use if you feel that something is wrong Always be careful to inspect this product for looseness rattling tilting wobbling sounds temperature odors etc Immediately...

Page 8: ...forcefully This could cause damage to covers or defective function 11 Be careful not to heat an empty water heater Exercise caution as heating of an empty water heater may result in burning of the hea...

Page 9: ...even after turning on the main switch Turn on the power of the compressor before operating this product 4 Use the turbine with a water check valve Use the turbine with a water check valve Contact the...

Page 10: ...to be trapped between the sub link cover and the base plate Do not allow feet or obstruct between the sub link cover and the base plate 2 3 Take care not to be trapped by moving parts of the headrest...

Page 11: ...thinners or oil based alcohol butanol and isopropyl alcohol which may cause cracks 4 Precautions for cleaning the operation panel membrane switches Penetration of droplets of sanitizing spray into the...

Page 12: ...Cup filler water heater Bowl flush Approximately five minutes for water changing in the cup filler line Handpiece line Pick up handpieces from the holder one at a time leave the turbine untouched and...

Page 13: ...ght Light Pole First Arm Horizontal Arm Balance Arm Doctor Unit Section Doctor Unit Section Cuspidor Unit Section Cuspidor Unit Section Foot Controller Light Pole Dental Light First Arm Horizontal Arm...

Page 14: ...ctor Unit Control Panel Rod Type Continental Type Instrument Holder Doctor Unit Control Panel Handle Sub Tray Rest Holder Cuspidor Unit Section Cupfiller Nozzle Bowl Flush Nozzle Water Tank Assistant...

Page 15: ...AC 230V In the junction box Protective Earth Seal Functional Earth Seal The Rated maximum supply pressure WATER 0 6MPa AIR 0 8MPa Seal for Fuse Seal Rated Maximum Supply Pressur F U S E F U S E T5AL...

Page 16: ...iately Handpiece Light ON OFF Switch Switching between on and off of the handpiece light occurs each time when this switch is changed Be sure to operate switches with your hands If operate with other...

Page 17: ...oves downward to the initial position and the backrest rises to upright position Press any switches chair unit for stopping automatic movement Switches provide manual operation for up down and backres...

Page 18: ...switch for about 5 seconds The buzzer sounds from the chair and setup has been completed 3 To change the set position perform the procedures steps 1 and 2 above LP Confirm that the patient is seated i...

Page 19: ...c Screen LED light source h 14 Your eyes could be dazzled by LED light not only under dark environment but also under normal environment Be careful about the brightness of the ambient Use of controls...

Page 20: ...OFF Switch The knob adjusts the output power of the electric scaler The output power increase when the knob is turned to clockwise and the output power decrease when the knob is turned to counterclock...

Page 21: ...OFF Switch The knob adjusts the output power of the electric scaler The output power increase when the knob is turned to clockwise and the output power decrease when the knob is turned to counterclock...

Page 22: ...The pedal depressing extent can control the turbine and air motor rotation speed Coolant water switch allows handpiece coolant water to be turned on or off Chip air for blowing off the chips jets out...

Page 23: ...heater may result in burning of the heater 2 Syringe air control Make adjustment of syringe air by the syringe air control knob Refer to ADJUSTMENT OF PARTS sections 3 Syringe water control Make adjus...

Page 24: ...ned to holder Opening or closing the slide knob can control the suction ow rate Water is supplied through from the cup ller nozzle when this switch is pressed Do not press the switch while the cup is...

Page 25: ...bottle may be used The water tank is intended only for use with purified water distilled water and pure water Do not use mouthwash or electrolyzed water as they may cause clogged tubing or affect int...

Page 26: ...n be stop by turning the shut off valve clockwise and open by turning the shut off valve counterclockwise Drain Valve Drain valve is used to discharge water from the air filter Chair motion stop funct...

Page 27: ...he flow rate but do not serve as stop valves Caution should be exercised as turning the knob excessively may cause it to loosen Service outlet water flow control 1 Cuspidor Unit Section 2 Flow Adjustm...

Page 28: ...ng so that the main air pressure gauge reads a value between 0 45 and 0 5 MPa Be sure to push the knob to lock it after setting This knob is adjusted when installing the unit Do not change the setting...

Page 29: ...oil mist separator collects oil contained in exhaust air from the handpieces Be sure to clean the oil mist filter once a month The oil mist filter can be detached by removing two screws Replace the oi...

Page 30: ...cleaning and disinfection Wipe off with dry soft cloth to dried up of the instrument holder after cleaning and disinfection Use FD333 or FD366 made by Durr or ethanol as sprayed to a soft cloth for c...

Page 31: ...table control panel after cleaning and disinfection When the surface of the operation panel is cleaned with disinfectant etc wipe off disinfectant completely If it penetrates into the back of the she...

Page 32: ...of the vacuum is reduced When the lid is removed and the bar is pulled the filter is also drawn out with the bar The filter only may be detached when the filter is turned The sucking unit comes into...

Page 33: ...Handpiece Saliva Ejector Handpiece Disassembly Disassemble the handpiece for the preparation of cleaning as the following gures show Pull the hose connector to disconnect the vacuum hose Cleaning by...

Page 34: ...aliva Ejector Handpiece can be autoclave Vacuum handpiece and saliva ejector handpiece have to assemble before autoclave When assembling the vacuum handpiece align the two surfaces a at surface of the...

Page 35: ...ed in an autoclave many times However this does not have a negative effect on performance If damage occurs to the sterilization pouch discard and sterilize again by using a new pouch Storage After ste...

Page 36: ...ir filter in the junction box needs to be replaced at least once every three years Contact your local service representative for replacement Water filter Sintered metal element Rated filtration 100 m...

Page 37: ...ing on the specifications AC230V 50Hz 60Hz 3 0A 2 8A 5A 250V Low Breaking Capacity Fuse Time lag 0 45 0 5 MPa 0 1 0 2 MPa 70kg without dental light 300LED Dental Light Type 320M Sensor Type CREDIA G1...

Page 38: ...lues are the standard values Unit mm Dimensional tolerance 10 Holder Type 760 367 650 1149 1024 1302 Rod Type Continental Type 760 367 294 650 1099 1302 490 522 384 811 686 70 70 70 70 1448 158 158 71...

Page 39: ...35 180 225 270 and 315 0 UT 1cycle and 70 UT 25 30 cycles at 0 single phase 0 UT 250 300 cycles 0 UT 0 5 cycles 0 45 90 135 180 225 270 and 315 0 UT 1cycle and 70 UT 25 30 cycles at 0 single phase 0 U...

Page 40: ...including peripherals such as antenna cablesand external antennas should be used no closer than 30 cm 12 inches to any part of the CREDIA G1 including cables specified by the manufacturer Otherwise d...

Page 41: ...es or air motors may not fit into some handpieces due to the manufacturing tolerances Have your local authorized Belmont dealer check the connectability before purchasing the handpiece Except for our...

Page 42: ...pon checkup stated above then stop using the unit turn off the main switch and contact your dealer or our of ce No water from handpiece Handpiece water adjustment knob is closed Open the knob Turbine...

Page 43: ...and electric accidents 2 Water main valve Be sure to turn the main water valve counterclockwise to the close position at the end of each work day 3 AIR main valve Be sure to turn the main air valve co...

Page 44: ...re may arise Control the water flow in accor dance with Control of compo nents described in the instruction manual If any other abnormality arises refer to the instruction manual attached to individua...

Page 45: ...scaler tip For each patient Appropriate tip shall be positively mounted and be correctly used Make sure to refer to the instruction manual attached to the scaler The tip will not normally work and tro...

Page 46: ...ctor handpiece body Foot controller Water supply hose Drain hose Air supply hose Electric wiring of moving part Standard Lifetime No 3 years 1 2 3 years 3 5 years 3 years 3 4 4 5 3 years No Parts Desc...

Page 47: ...t Male connector for water service outlet Intended use Optional Accessories 42 The container to take in cotton and swab for dental treatment Receptacle to do away with used cotton etc Viewer for X ray...

Page 48: ...NTAL UNIT CLASS IIa Model CREDIA G1 CLASS IIa has been de ned by the rule 9 of MDD Annex IX The product has been designed and manufactured in accordance with the European standards as the listed in th...

Page 49: ...NT CORPORATION BELMONT MANUFACTURING CO LTD Manufacturer TEL 84 2513 201 100 FAX 84 2513 201 096 Lot I 2 Long Duc Industrial Park Long Duc Ward Long Thanh District Dong Nai Province Vietnam TAKARA COM...