Unit part
75
Doctor’s unit
77 type 3-way syringe
Disassembly
Remove the nozzle from syringe by turning it in direction A.
Cleaning by hand
・
Wipe off the surface contamination by a cloth while rinsing
the surface by running clean warm water at 40±5 degrees.
Scrub the tip and joint part of nozzle by a cleaning brush or
by a tooth brush with running clean warm water at 40±5
degrees.
・
Rinse thoroughly by distilled water at ordinary temperature
or by clean water for more than 1 minute.
・
Check whether contamination is removed or not after
cleaning.
Continue the cleaning if contamination is remained.
・
Immersed with an alkaline detergent for 5 minutes.
(We recommend to use ID212 made by DURR)
・
Rinse thoroughly by distilled water at ordinary temperature
or by clean water for more than 1 minute.
Sterilization
The nozzle can be sterilized with autoclave. Sterilization with
autoclave is permitted up to 250 times.
・
Insert the handpiece in a sterilization pouch and seal it.
・
Autoclave for 3 min. at 134
℃
and dry for 15min..
Apply Vaseline thinly and evenly to the two O-rings after
sterilization.
Tighten the nut firmly in direction B to undo the nozzle.
Storage
After sterilizing the nozzle, keep it in the sterilization pouch
and store in a dark and cool place.
CAUTION
Cleaning must be done every after use to patients.
Cleaning must be done within 1 hour after use. Replace with a new syringe nozzle if contamination and
solid material attached to syringe nozzle can not be removed.
CAUTION
• Sterilization must be done every after use to patients.
• Sterilization by class B cycles.
• Sterilization temperature is 135 degrees or less.
• Dry naturally if the temperature for drying process is to exceed 135
℃
.
• If damage occurs to the sterilization pouch, discard and sterilize again using a new pouch.
• Before use, make sure that the nut is firmly tightened.
NOTICE
For operation of other syringes, observe the instructions shown in the package inserts and Instruction
Manuals included with the syringes.
CLEANING
Summary of Contents for CP-ONE PLUS
Page 12: ...Chair part 8 OVERVIEW AND MAJOR COMPONENTS Seat Legrest Roll up Legrest...
Page 24: ...Unit part 20 Base mount type Cart type OVERVIEW AND MAJOR COMPONENTS...
Page 25: ...Unit part 21 OVERVIEW AND MAJOR COMPONENTS...
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ...NOTE BOOK NO 1A07NMME Printed in Japan 2017 09...