Unit part
57
Function switch
Precautions for setting water feeding quantity
▪ Carry out setting in a condition where water comes out of the cupfiller nozzle normally. Water may not
be fed as set if setting is carried out in a condition where water contains air, etc.
▪ If
is pressed during setting, setting done till that time is invalidated and the initial status
immediately after starting setting is restored. (The buzzer sound changes to “beep, beep”.)
▪ The water feeding quantity may be less than the set quantity, if “Cup/linked with bowl” is used.
▪ If the flow rate (water pressure) of water in use changes after setting, the water feeding quantity may
change.
* If the cup water feed control valve is adjusted, the flow rate changes and accordingly the water feeding
quantity may change.
▪ Set the quantity so that the total weight of the cup and water is 200 g or less.
(Reference) Paper cup for dental use (approximately 3.5 g): The quantity may be set to a cup of water.
Stainless steel cup for dental use (approximately 60 g): The quantity may be set to approx. 80% of the
cup (approximately 140 cc).
* Water may not be fed as set if the total weight exceeds 200 g.
▪ Be sure to set the water feeding quantity, if the type of cup to be used is changed (e.g., paper cup is
changed to stainless steel cup).
▪ Be sure to set the water feeding quantity, if the flow rate of feed water is changed with the cup water
feed adjustment knob.
▪ An error may occur and the buzzer near the cup table may sound after setting, if the water feeding
quantity is not set properly. Retry the setting of the water feeding quantity from the beginning, if the
buzzer sounds.
▪ Set the water feed time between 1 and 12 seconds. If the water feed time is less than 1 second, an error
may occur and the buzzer may sound after completion of setting. Retry the setting of the water feeding
quantity from the beginning, if the buzzer sounds. Water feeding is disabled and the water feed time is
set to 12 seconds, if the water feed time exceeds 12 seconds during setting.
▪ Set the water feeding quantity over 20 cc. The buzzer may sound after setting if the water feeding
quantity is less than 20 cc. Retry the setting of the water feeding quantity from the beginning, if the
buzzer sounds.
(iii) On/Off setting for interlocking operation
between the cup filler and the bowl rinse
Individual mode
The spittoon bowl is not cleaned when cup water
feeding starts.
Link mode
The spittoon bowl is cleaned as soon as cup water
feeding starts.
[Setting procedures]
1. Select “CUP/BOWL LINK” from the functions.
2. Select intended setting with
or
and press
to save.
CUP/BOWL LINK
LINK
INDIVIDUAL
Link
OPERATION SETTING
Summary of Contents for CP-ONE PLUS
Page 12: ...Chair part 8 OVERVIEW AND MAJOR COMPONENTS Seat Legrest Roll up Legrest...
Page 24: ...Unit part 20 Base mount type Cart type OVERVIEW AND MAJOR COMPONENTS...
Page 25: ...Unit part 21 OVERVIEW AND MAJOR COMPONENTS...
Page 93: ......
Page 94: ......
Page 95: ......
Page 96: ...NOTE BOOK NO 1A07NMME Printed in Japan 2017 09...