Belmont Clesta II Operating Instructions Manual Download Page 32

6. MAINTANANCE AND INSPECTION

6-1. Guide for daily maintenance and inspection (Maintenance and inspection by user)

 •Management of maintenance and inspection of medical equipment should be implemented by the user 

    (medical institution). 
 • For safe use of this product, it is necessary that inspection 

should be conducted in the specified frequency

    on the tems described below.

Air turbine revolution, 

water flow, air flow and so 

forth shall be free of 

abnormality.

Micromotor revolution, 

water flow and so forth shall 

be free of abnormality.

Scaler vibration, water 

flow and so forth shall be 

free of abnormality.

No.     Item        Frequency  Inspection method 

and diagnosis 

Influence if inspection 

not conducted

Maintenance required in case 

of nonconformity 

Check

of safety 

functions

Before

start

Before

start

Before

start

Before

start

For each 

patient

For each 

patient

For each 

patient

After

closing

Unexpected personal 

injury and troubles may 

arise due to motion of

the chair during medical

treatment and due to

pinching between 

doctor section and 

chair.

Contact your dealer or our 

office if any abnormality arises.

Check for 

leakage of 

water and 

air

Leakage of water and air 

shall not be observed 

around the product. 

The product will not

normally work, and 

troubles may arise. 

Contact your dealer or our office

if any abnormality arises.

Cupfiller

When a paper cup is placed 

on the cupfiller, the cup 

shall be detected and 

cupfilling shall be executed.

* Malfunction may arise if 

the cup is of another 

material grade (such as 

stainless steel and plastics) 

or if the paper cup is of dark 

color or pattern.

Cupfilling may not be 

executed.

Execute re-inspection in accor-

dance with "Method for operation" 

described in the instruction

manual.

Contact your dealer or our office if 

recovery is not achieved as a 

result of re-inspection. 

Check of

motions of

equipment

Troubles such as injury 

in patient's oral cavity 

and equipment failure

may arise. 

Control the water flow in accor-

dance with "Control of compo-

nents" described in the instruction

manual.  If any other abnormality 

arises, refer to the instruction

manual attached to individual 

equipment.

Contact your dealer or our office if 

recovery is not achieved.

Check of 

air turbine 

bar

Appropriate bar shall be 

positively mounted.

Make sure to refer to the

instruction manual attached

to individual equipment. 

The bar will not normally 

work and troubles may 

arise.

If abnormality such as flaw and 

deformation is found on the bar,

replace the bar in accordance with 

the instruction manual attached to

individual equipment. 

Check of

scaler tip

Appropriate tip shall be 

positively mounted and be 

correctly used.

Make sure to refer to the 

instruction manual attached 

to the scaler.

The tip will not normally 

work and troubles may 

arise.

If the tip was worn or deformed, 

replace the tip in accordance with 

the instruction manual attached to

the scaler.

Contact your dealer or our office if 

any other trouble arises.

Check of

tightness of

syringe nut

The nut for fixing the nut of

Type 77, 3-way syringe shall 

be positively tightened.

Troubles may arise if

the nut comes off.

Turn and positively retighten the 

nut that fixes the nozzle. 

Matters

attached to 

micromotor

Excessive handpiece oil or

the like shall not be attached 

to the motor section.

The motor section will 

not work normally and 

troubles may arise. 

Execute care in accordance with 

the instruction manual attached to

individual micromotor.

Make sure the chair 

movement stops by any of 

the following actions.

when foot controller pedal 

is depressed. 

While setup is in progress 

with function switch on the 

doctor membrane switch panel.

When the spittoon bowl is 

turned to patient side 

(Pedestal type) 

-25-

  

During chair auto

movement , press for any

chair operation switch. 

Summary of Contents for Clesta II

Page 1: ...before operating the unit After the installation is completed le this manual and refer back to it for future maintenance 0197 If you have any questions about this Manual or this product please contac...

Page 2: ......

Page 3: ...N 18 4 SAFETY LOCK DEVICE 18 5 CARE AND MAINTENANCE 19 6 MAINTANANCE AND INSPECTION 6 1 GUIDELINE FOR DAILY MAINTENANCE AND INSPECTION 25 MAINTENANCE AND INSPECTION BY USER 6 2 GUIDELINE FOR PERIODICA...

Page 4: ......

Page 5: ...dental light Handpieces that can be used with this product Use the Handpieces described on page 30 Chair that can be used with this product Use the Chair described on page 30 Dental Light that can be...

Page 6: ...s SYMBOLS In this manual on the labels or on the control panel of CLESTA ROD TYPE following symbols are used Confirm the meaning of each symbol Serial number Reference number 1 The color of black part...

Page 7: ...lush About 5 minutes for cleaning water in cupfiller line Flushout procedure Handpiece line Pick up one handpiece out of the holder and hold it over the cuspidor bowl with the turbine kept as it is an...

Page 8: ...turn off the main switch upon completion of work or during work breaks This prevents incorrect operation due to accidental contact and associated hazards 4 Do not place an undue load on the arm Do no...

Page 9: ...ients with a cardiac pacemaker Use this product with extreme caution on patients with a cardiac pacemaker In the case of any abnormalities in patients during use immediately turn off this product and...

Page 10: ...sensor may react to hands Resulting water spill 12 Immediately wipe off drug solution when it comes into contact with this unit Should drug solution or water comes into contact with this unit immedia...

Page 11: ...of ush out Practice the ush out of water retained in the unit before the start of clinical practice at the beginning of each work day to maintain the quality of water for dental treatment and ensure...

Page 12: ...r check valve The water unit is intended for use with tap water Caution should be exercised as the use of water other than tap water water through a sterilizer of water systems etc may result in failu...

Page 13: ...andle Right Hand Side Standard Left Hand Side Option Rod Dental Size Film Viewer Handpiece Rest Sub Tray Holder Foot Controller Dental Light Handpieces Micromotor Air Turbine Motor Scaler and etc Manu...

Page 14: ...700 21 28 890 274 9 0 517 9 0 90 1 3 8 441 184 569 60 60 5 651 489 737 50 8 42 7 600 750 187 723 294 666 194 90 549 723 294 666 194 90 549 Fig 2 1 Dimensions Chair Mount Type Pedestal Type 7 10 to 95...

Page 15: ...icator MAIN LOCK 1 2 0 LP 2 1 6 4 3 7 5 MAIN LOCK 1 2 0 LP F 2 1 4 3 5 16 15 13 11 7 14 6 9 10 12 8 Cup ller Switch Fig 3 5 Momentarily press the cup ller switch water will come out from the cup ller...

Page 16: ...outh rinsing position and stops automatically Momentarily press LP switch again the backrest returns to the previous treatment position and stops automatically d Emergency Stop Safety Stop During auto...

Page 17: ...the xed rotation speed indicated on the indicator Dental Light ON OFF Switch Fig 3 14 Switch for on off the dental light Fig 3 12 Electric Motor Speed Set Switch Fig 3 13 Electric Motor Rotation Direc...

Page 18: ...memory 0 Switch 1 Switch 2 Switch LP Switch 0 1 2 2 Preset time setting Store Switch Function Switch Decrease Switch Increase Switch 3 Preset time operation 4 Cancel the timer during time countdown 0...

Page 19: ...times and momentarily press the increase switch Function Switch Decrease Switch Increase Switch 1 Short time ush out Momentarily press the function switch three times and momentarily press the decreas...

Page 20: ...ctronic sound for timer Electronic sound for timer can be changed Momentarily press the function switch six times Momentarily press one of chair auto mode switch 0 1 2 LP then the new electronic sound...

Page 21: ...drive air pressure is shown in the pressure gauge Fig 3 16 Doctor Table Section Control Knobs 5 Balance Arm Air Brake Fig 3 17 When the master switch is ON the balance arm is locked by the air brake...

Page 22: ...witch is located on the cuspidor unit control panel Turn on the water heater switch and the cup ller water will warm up C Service Water Outlet Fig 3 19 The service water outlet provides a quick connec...

Page 23: ...ls up the cup and stops automatically When cup ller starts the bowl ush also starts and will run for about 6 sec and it stops automatically While lling the cup momentarily press the cup ller switch an...

Page 24: ...or electrolyzed water as they may cause clogged tubing or affect internal valves and equipment Adjust the air supply pressure for the water tank to 200 kPa or less An excessively high pressure may ca...

Page 25: ...pedal further to the right The speed control by the foot control is within the limits of the electric motor speed setting E Coolant Water Switch Electric Motor Rotation Direction Switch Optional The...

Page 26: ...Bowl A Detach the drain cap and clean the basket strainer Remove the cup ller base and clean it Turn the bowl ush nozzle counterclockwise avoiding the clash with the cuspidor bowl Fig 5 1 B Turn the c...

Page 27: ...clean and sterilize it at the end of each work day Recommended cleaner Orotol Plus made by DURR 5 Vacuum hose and saliva ejector hose are detachable from the cuspidor unit Fig 5 6 Turn the hose conne...

Page 28: ...each Fig 5 10 B Check whether contamination is removed or not after cleaning Continue the cleaning if contamination is remained C Immersed with an alkaline disinfection or detergent for 5 minutes We r...

Page 29: ...Insert the handpiece in a sterilization pouch and seal it B Autoclave for 3 min at 134 and dry for 15 min Sterilization with autoclave is pemitted up to 250 times Fig 5 12 Vacuum Handpiece and Saliva...

Page 30: ...ntamination is removed or not after cleaning Continue the cleaning if contamination is remained C Immersed with an alkaline disinfection or detergent for 5 minutes We recommend to use ID212 made by DU...

Page 31: ...ig 5 17 The water lter in the junction box needs to be replaced at least once a year The air lter in the junction box needs to be replaced at least once every three years Contact your local service re...

Page 32: ...chieved as a result of re inspection Check of motions of equipment Troubles such as injury in patient s oral cavity and equipment failure may arise Control the water flow in accor dance with Control o...

Page 33: ...h cannot be turned off or if the main valve cannot be closed Product s moving parts No abnormal noise or the like shall be produced from product s moving parts when the product is operated The product...

Page 34: ...tion Standard Lifetime No 3 years 8 9 3 years 10 5 years 5 years 11 12 3 years Regulator Valves Switches Film viewer body part Pressure gauge Arm section of moving part Control PCBs 6 3 years 13 7 yea...

Page 35: ...45 90 135 180 225 270 and 315 0 UT 1cycle and 70 UT 25 30 cycles at 0 single phase 0 UT 250 300 cycles 0 UT 0 5 cycles 0 45 90 135 180 225 270 and 315 0 UT 1cycle and 70 UT 25 30 cycles at 0 single ph...

Page 36: ...ls such as antenna cablesand external antennas should be used no closer than 30 cm 12 inches to any part of the CLESTA including cables specified by the manufacturer Otherwise degradation of the perfo...

Page 37: ...ce hose are designed and manufactured in accordance with ISO9168 2009 DENTISTRY HOSE CONNECTORS FOR AIR DRIVEN DENTAL HANDPIECES However there still could be a case that the connectors of turbines or...

Page 38: ...er to take in cotton and swab for dental treatment Receptacle to do away with used cotton etc intended use etc Model number MCL 04NH 1B Quick connector that provides external equipment with air Model...

Page 39: ...I Product DENTAL UNIT CLASS IIa Model CLESTA CLASS IIa has been de ned by the rule 9 of MDD Annex IX The product has been designed and manufactured in accordance with the European standards as the li...

Page 40: ...42 0083 Japan TEL 81 6 6213 5945 FAX 81 6 6212 3680 TAKARA BELMONT CORPORATION Printed in Japan 2019 12 BOOK NO FEFA27Y0 NOTE TAKARA COMPANY EUROPE GmbH Berner Strasse 18 60437 Frankfurt am Main Germa...

Reviews: