-23-
(7) Belmont 77 Syringe
* Cleaning amd sterilization must be done every after use
to patients.
Disassembly (Fig.5-13)
Fig.5-13
Fig.5-14
Fig.5-15
Fig.5-15
Remove the nozzle from syringe by turning it in direction A.
Cleaning by hand
A. Wipe off the surface contamination by a cloth while rinsing
the surface by running clean warm water at 40±5 degrees.
Scrub the tip and joint part of nozzle by a cleaning brush
or by a tooth brush with running clean warm water at 40±5
degrees.
(Fig.5-14)
B. Check whether contamination is removed or not after
cleaning. Continue the cleaning if contamination is
remained.
C. Immersed with an alkaline disinfection or detergent for 5
minutes. (We recommend to use ID212 made by DURR)
D. Rinse thoroughly by distilled water at ordinary temperature
or by clean water for more than 1 minute.
• Sterilization by class B cycles.
• Sterilization temperature is
135℃ o
r less.
• Dry naturally if the temperature for drying process is to exceed 135
℃
.
• If damage occurs to the Sterilization pouch, discard and sterilize again using a new
pouch.
• Before use, make sur
e that the nut is firmly tightened.
Storage
After sterilizing the nozzle, keep it in the sterilization
pouch and store in a dark and cool place.