Function Switch (Fig.3-15)
Use this switch for setting various working conditions.
As for the setup procedures, please refer to 3-1.(3) a-i.
12
(3) Function Swicth Setup Procedure
1) To set a timer
Momentarily press the function switch, and set the time by pressing decrease switch and increase switch.
… Minimum setting time by switch is 10 seconds.
… Minimum setting time by switch is 1 minute.
The setting time is indicated on the function indicator.
Momentarily press the start switch to start timer. The end of setting time is informed by electronic sounds.
Example: Setting time 3 minutes 30 seconds is indicated as 03:30 in the function indicator.
a. Timer
Timer can be set maximum 90 mins. 50 secs. in 10 secs. segment.
Function Switch
Decrease Switch
Increase Switch
+
Start Switch
(Electric Motor Speed Set Switch)
+
Four preset time can be set. (0) (1) (2) (LP)
Momentarily press function switch, and set the time by pressing decrease switch and increase switch.
Press the store switch, then press the (0)(1)(2) or (LP) switch to store in memory.
0 Switch 1 Switch 2 Switch LP Switch
0
1
2
2) Preset time setting
Store Switch
Function Switch Decrease Switch Increase Switch
+
3) Preset time operation
4) Cancel the timer during time countdown
0 Switch 1 Switch 2 Switch LP Switch
0
1
2
LP
Function Switch
Start Switch
(Electric Motor Speed Set Switch)
Press the function switch, then press the (0)(1)(2) or (LP) switch to choose desired preset number.
Press start switch to start timer.
Momentarily press function switch, then press start switch to cancel timer.
Function Switch
Start Switch (Electric Motor Speed Set Switch)
E
A
E
A
Fig.3-15 Function Switch
b. Group Selection Mode
The group selection mode is a convenient function for the dental clinic where a multiple (up to four groups)
of dentists work with one unit. The following functions can be set for each group.
Preset position for chair. / Preset rotation speed for micro motor.
To set the group.
1) Momentarily press the function switch twice on main control panel, and the function indicator will
indicate group number.
2) Momentarily press the 0,1, 2 or LP switch on main control panel to set one of 4 group.
0 ; Group1 / 1 ; Group2 / 2 ; Group3 / LP ; Group4
0
1
2
LP
Function Switch
0 Switch
1 Switch
2 Switch
LP Switch
LP
-11-