SAFETY PRECAUTIONS
CAUTION
18. Precautions for cleaning the operation panel (membrane switches)
The water tank is intended only for use with purified wate
r, distilled water and pure water. Do not use mouthwash
or electrolyzed water, such as ConCool or povidone iodine, as they may cause clogged tubing or affect internal
valves and equipment.
23. Read the documents accompanying the various pieces of equipment
Before use, be sure to carefully read the package inserts and Instruction Manuals accompanying the various
pieces of equipment (including optional articles) to ensure proper use.
Penetration of droplets of sanitizing spray into the back of the operation panel may be associated with switch failure.
Use a paper towel soaked with sanitizing solution to clean the surface of the operation panel.
21. Set the pressure of the water tank at 200 kPa or less
Adjust the air supply pressure for the water tank to 200 kPa or less. An excessively high pressure may cause
damage to the water tank.
19. Precautions for cleaning
20. Practice of flush out
Practice the
fl
ush out of water retained in the unit before the start of clinical practice at the beginning of each
work day to maintain the quality of water for dental treatment and ensure a steady supply of water to handpieces.
Refer to the procedure for flush out.
22. Do not use water other than purified wate
r, distilled water or pure water for the water tank
• Never use sandpaper, metal scrub brushes and abrasive cleaning agents to clean the unit.
-4-
24. Precautions for sterilization
Do not sterilize other than specified method. It may cause device failure.
27. Precautions for sterilization
Do not sterilize other than specified method. It may cause device failure.
• Do not use acidic cleaning agents or alkaline pipe cleaning agents to avoid corrosion of metal and resin parts,
etc. Never use cleaning agents that include sodium hypochlorite.
25. Combination with accessories or other devices
Use only our qualified accessories, other devices or equipment fo
r this product to ensure the safety of product.
26. Pay attention to allergic reaction of a patient
While a vaccume tip is placed in a patient mouse, pay attention to allergic reaction.
If allergic reaction start to happen, immediately stop the usage of a vaccume tip.