2 Safety Consideration
2-3 EMC Information
This product complies with EMC Standard EN60601-1-2:2015.
1. Precautions regarding EMC and compliance with
accompanying documents
Medical electrical equipment requires special precautions
regarding EMC and needs to be installed and put into service
according to the EMC information provided in this booklet.
2. Effects of RF communication devices
Portable and mobile RF communication devices can affect
medical electrical equipment.
3. Installation exclusion environment
Hospitals except for near active HF SURGICAL
EQUIPMENT and the RF shielded room of an ME SYSTEM
for magnetic resonance imaging,where the intensity of EM
DISTURBANCES is high.
4. Electromagnetic emission declaration
Group 1
Emissions test
Compliance
Electromagnetic environment—guidance
Class B
Guidance and manufacturer’s declaration—electromagnetic emissions
This product is intended for use in the electromagnetic environment specified below.
The customer or user of this product should ensure that it is used in such an
environment.
This product only uses RF energy for its
internal functions.Therefore, its RF emissions
are very low and are not likely to cause any
interference with nearby electronic equipment.
This product is suitable for use in all establish-
ments, including domestic establishments and
those directly connected to the public
low-voltage power supply network that
supplies buildings used for domestic purpos-
es.
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC 61000-3-2
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Class A
Complies
Use of this equipment adjacent to or stacked with other equipment
should be avoided because it could result in improper operation.
If such use is necessary,this equipment and the other equipment
should be observed to verify that they are operating normally.
WARNING