[ 13 ] ELECTROMAGNETIC COMPATIBILITY (EMC)

 

 

This product conforms to EMC standard IEC60601-1-2:2014.

 

 

1. Caution to EMC and Compliance with information in attached document

 

 

Medical electrical equipment requires special attention to EMC and it must be installed and used 

according to the EMC information provided in this instruction manual. Do not install in the 

vicinity of the electrosurgical device being output or electromagnetically shielded room of ME 

system for MRI diagnostic imaging because the electromagnetic interference intensity is high.

 

 

WARNING

 

a. Use of this equipment adjacent to or stocked with other equipment should be avoided 

because it should result in improper operation.  If such use is necessary, this equipment 

and the other equipment should be observed to verify that they are operating normally.

 

 

b. Use of accessories, transducers and cables other than those specified or provided by the 

manufacturer of this equipment could result in increased electromagnetic emissions or 

decreased electromagnetic immunity of this equipment and result in improper operation.

 

 

c. Portable RF communications equipment (including peripherals such as antenna cables and 

external antennas) should be used no closer than 30cm (12 inches) to any part of the

 

BELRAY II MODEL 097, including cables specified by the manufacturer.  

Otherwise, degradation of the performance of this equipment could result.

 

 

 

2. Electromagnetic emissions

 

 

 

Emissions test

 

 

Test procedure

 

 

Compliance

 

Note :

 

The emissions characteristics of this equipment make 
it suitable for use in industrial areas and hospitals 
(CISPR 11 class A). If it is used in a residential 
environment (for which CISPR 11 class B is normally 
required) this equipment might not offer adequate 
protection to radio-frequency communication 
services.  The user might need to take mitigation 
measures, such as relocating or re- orienting the 
equipment.

 

Conducted and 

radiated RF 

emissions

 

 

CISPR11

 

 

Group 1 Class A

 

Harmonic distortion

 

 

IEC61000-3-2

 

 

Class A

 

 

Voltage fluctuations 

and flicker

 

 

IEC61000-3-3

 

 

Clause 5

 

 

3. Electromagnetic immunity

 

 

 

Immunity test

 

IEC60601 

test level

 

Compliance level

 

Electromagnetic environment - 

guidance

 

Electrostatic 

discharge (ESD) 

IEC61000-4-2

 

±8 kV contact

 

±15 kV air

 

±8 kV contact

 

±15 kV air

 

Floors should be wood, concrete or 

ceramic tile. If floors are covered 

with synthetic material, the relative 

humidity should be at least 30%.

 

Electrical fast 

transient/burst 

IEC61000-4-4

 

±2 kV for

         

power supply lines

 

±1 kV for 

input/output lines

 

±2 kV for

         

power supply lines

 

±1 kV for 

input/output lines

 

Mains power quality should be that 

of a typical commercial or hospital 

environment.

 

Surge

 

IEC61000-4-5

 

±1 kV differential mode

 

±2 kV common mode

 

±1 kV differential mode

 

±2 kV common mode

 

Mains power quality should be that 

of a typical commercial or hospital 

environment.

 

Power frequency 

(50/60Hz) 

magnetic field

 

IEC61000-4-8

 

30 A/m

 

30 A/m

 

Power frequency magnetic fields 

should be at levels characteristic of 

a typical location in a typical 

commercial or hospital 

environment.

 

 

- 12 - 

Summary of Contents for BELRAY II 097

Page 1: ...R OPERATOR S INSTRUCTIONS for USA Canada WARNING This X ray unit may be dangerous to patient and operator unless safe exposure factors operating instructions and maintenance schedules are observed DEN...

Page 2: ...F CONTROLS 4 4 OPERATING PROCEDURES 7 5 ESTIMATED AIR KERMA 7 6 OPTIONAL HAND EXPOSURE SWITCH 7 7 DIGITAL IMAGING SYSTEM 8 8 DISINFECTION AND CLEANING 8 9 ERROR CODES 9 10 MAINTENANCE 10 11 DISPOSAL 1...

Page 3: ...housing assembly 097 H b X ray controls 097 CM main controller 097 CS sub controller c Cones 097 R regular 097 L long d Balance arm 097 A 4 COMPLIANCE WITH STANDARD The BELMONT BELRAY II MODEL 097 x r...

Page 4: ...re warning light be at least 2 m away from the x ray head and patient and out of the path of the x ray beam or be positioned behind a protective device fully use all radiation protection devices acces...

Page 5: ...justing Switch Down 12 Digital Imaging Switch 4 ExposureTimeAdjusting Switch Up 13 4 mA Selction Switch 5 Tooth Selection Switch T1 6 Tooth Selection Switch T2 7 Tooth Selection Switch T3 8 Tooth Sele...

Page 6: ...I 0DQGLEOH 7 QFLVRU RI 0D LOOD XVSLG 3UHPRODU RI 0DQGLEOH 7 XVSLG 3UHPRODU RI 0D LOOD 0RODUV RI 0DQGLEOH LWHZLQJ 7 0RODU RI 0D LOOD LWHZLQJ 0RODUV 7 2FFOXVDO I WKH 7 VZLWFK LV GHSUHVVHG PRUH WKDQ VHF...

Page 7: ...PRPHQWDULO GLVSOD V WKH VSHHG EHLQJ VHOHFWHG LQ WKH SRVXUH 7LPH LVSOD LQGRZ LWK WKH IDFWRU VSHHG VHWWLQJ G 08 WKH H SRVXUH WLPH EHFRPHV KDOI RI 08 VHWWLQJ As the sensitivity is different according to...

Page 8: ...osure and an error code E 00 will be displayed in Exposure Time Display Window 16 6 switch once after setting to child If the weight of child is over 30kg and less than 50kg press Speed Setting mA Chi...

Page 9: ...6ZLWFK in RUGHU WR SUHYHQW DFFLGHQWDO exposures when the unit is not in XVH 127 I WKH XQLW OHIW RYHU PLQ ZLWKRXW EHLQJ RSHUDWHG DQG WKH 0DLQ 3RZHU 6ZLWFK LV NHSW RQ ILJXUH UXQV WKURXJK WKH SRVXUH 7LPH...

Page 10: ...QLQJ RI WKH SDUWV WKDW PD FRPH LQ FRQWDFW ZLWK VNLQ SHULRGLF GLVLQIHFWLRQ ZLWK D QRQ FRUURVLYH VXUIDFH GLVLQIHFWDQW LV UHFRPPHQGHG 5HFRPPHQGHG GLVLQIHFWDQW XUU HQWDO 237 0 7 6FL DQ WG U LQJ OORZ VXUID...

Page 11: ...Malfunction of the microcomputer Tube current is detected during pre heating period Tube current is detected when main power switch is turned on Exposure switch or exposure circuit had been ON when ma...

Page 12: ...ed service technician The specific instructions to perform these checks are located within the BELRAY II MODEL 097 Installation Manual A Line voltage confirmation B Tube current confirmation C Inspect...

Page 13: ...long cone 1 9 3 3 mGy sec 40 Data obtained by direct measurement in the useful beam 16 Nominal electrical output of H V Generator 0 36kW at 70kVp 7mA 17 Cone Source to skin distance Field size a Regu...

Page 14: ...netic emissions Emissions test Test procedure Compliance Note The emissions characteristics of this equipment make it suitable for use in industrial areas and hospitals CISPR 11 class A If it is used...

Page 15: ...ding cables specified by the manufacturer Otherwise degradation of the performance of this equipment could result Radiated RF IEC61000 4 3 80 MHz 2700 MHz 3V m unmodulated r m s 80 AM 1kHz 80 MHz 2700...

Page 16: ...w Jersey 08873 U S A TEL 732 469 5000 800 223 1192 Fax 732 526 6322 800 280 7504 www belmontequip com TAKARA COMPANY CANADA LTD 2455 Meadowvale Boulevard Mississauga Ontario L5N 5S2 Canada TEL 905 816...

Reviews: