- 6 -

4. Care and Maintenance
4-1. Cleaning and Disinfection

Routine Care

Clean plastic and upholstery surfaces regularly using a mild soap and water solution. 

When cleaning, use a dampened cloth only,

 as excess cleaning solution can flow into the chair and cause

 perma- 

nent damage to electrical components. 

Barrier Technique

Use of disposable barrier products should be the first choice for the protection of dental equipment. Disinfectants 

leave behind a surface residue that accumulates over time and eventually damages equipment and upholstery

surfaces.

Chemical Disinfection 

If a chemical disinfectant is to be used on chair or upholstery surfaces, contact the manufacturer of the disinfec-

tant prior to use. 

 Obtain verification from the disinfectant manufacturer that their product will not damage the chair or upholstery

surfaces.

Unacceptable Disinfectants 

The following chemicals may damage equipment and upholstery:

* Alcohol based solutions 

* Bleach 

* Phenol / Alcohol combinations 

* Foam spray products 

* Benzalkonium chloride solutions

Use with Caution

* Water based phenolic disinfectants, following manufacturer’s exact instructions for use.

Warranty does not cover damage to equipment and upholstery caused by disinfectant solutions. 

NOTE

4-2-1 Storage

1. Keep the equipment away from 

standing water and puddles.

2. Keep in a circumstances safe from influence by 

extremes in temperature and humidity, and prevent excessive 

    exposure to wind,  sunlight, air containing salts and minerals.
3. Store product upright and level in a stable environment free from vibration. 

 

.

4. Do not keep the equipment in a place where chemicals are or where gas is emitted.

4-2-2 Lifetime

Provided that the recommended and authorized care and maintenance are correctly performed, the working life-

time of this equipment is 10 years from the date of initial shipment from the manufacturer.

4-2 Storage and Working Lifetime

CAUTION

Do not spray liquids directly onto chair surfaces and operation panel. 

In order to prevent damage to electrical components and systems, do not apply excess cleaning solution onto 

chair surfaces.

Summary of Contents for B50N

Page 1: ...manual provides operating instructions for the B50N Chair The instructions contained in this booklet should be throughly read and understood before operating the chair After the installation has been...

Page 2: ...rical Pressure 10psi 15 4psi 700hpa 1060hpa Storage Ambient Temperature 4 F 158 F 20 C 70 C Humidity 10 95 Atmospherical Pressure 10psi 15 4psi 700hpa 1060hpa Classi cation a Protection against electr...

Page 3: ...nt has been positioned Take extreme care with patients that have a pacemaker Prior to operation of the chair ask patient to verify whether they have a pacemaker When operating chair be certain patient...

Page 4: ...oot Control Ass y 17 Sub Link Cover Lower 18 Main Power Switch 19 Fuse Holder 20 Backrest Control Panels 21 Side Frame Left 22 Side Frame Right Figure 1 Figure 2 BACKREST CONTROL Figure 3 FOOT CONTROL...

Page 5: ...5 2 20mm Hospital grade plug Base component housing formed from solid color acrylate styrene acrylonitrile ASA V0 All exposed ferrous parts covered with corrosion resistant paint or plating Weight 32...

Page 6: ...of forward facing armrests or on their lap All patients are at risk of potential injury if these precautions are not followed Do not allow patient to sit on the headrest armrest backrest or on the fr...

Page 7: ...Pre set 3 is operated by LP button Auto Return Momentarily depress A button on the backrest control or 0 button on the foot control the chair will return to the initial position The seat is fully low...

Page 8: ...n so it is even with the occipital of the patient s head Viewing the lower arch 1 Using the latch release button white pivot the headrest cushion half way toward the patient s head 2 Have the patient...

Page 9: ...Bleach Phenol Alcohol combinations Foam spray products Benzalkonium chloride solutions Use with Caution Water based phenolic disinfectants following manufacturer s exact instructions for use Warranty...

Page 10: ...seat motion activation resulting in potential injury and or damage to equipment If the main power cannot be turned off discontinue use of product immediately and contact an authorized Belmont dealer f...

Page 11: ...i t p i r c s e D s t r a P o N s r a e y 2 m s i n a h c e M t s e r d a e H 1 2 Electric wiring of moving parts 2 years 3 Switches for operating equipment 2 years Transformer 3 years 4 3 Base Cover...

Page 12: ...sion of Takara Belmont USA Inc 101 Belmont Drive Somerset New Jersey 08873 U S A TEL 732 469 5000 800 223 1192 FAX 732 356 1035 Importer for Canada TAKARA COMPANY CANADA LTD 2455 Meadowvale Blvd Missi...

Reviews: