23
Start:
set start frequency of the digital sine wave generator. In order to avoid
overlapping frequencies, the start frequency may not be set to any value
above the stop frequency. Range: 50Hz - 10kHz.
Stop:
set stop frequency of the digital sine wave generator. In order to avoid
overlapping frequencies, the stop frequency may not be set to any value
below the start frequency. Range: 50Hz - 10kHz.
Test:
Starts frequency sweep within the set paramenters
User command: YES / NO .
Page 3
This function is identical to that of page 2, however it serves to check CH.B
Page 4
Parameters:
Sine Test:
Activates the automatic impedance check.
(as in funcions and features guideline p.3)
User command: ON / OFF
Power CTRL:
Activates the „Auto Power Off“ function
Re.: Protect LED at front
(as in funcions and features guideline p.3)
User command: ON / OFF
Remote Control*: Enables / disables remote control via the „Remote Active“ switch.
Re.: operational features (front): “Remote Active“ Switch.
User command: ENABLE / DISABLE .
*(function disabled in software release 1.0)
.