8
Selecting the Measurement Range
The HTR-101 has a 51” measurement span with which you may select the
upper and lower measurement values. The only limitation is the upper value
can not exceed 90” or 7ft 6in.
Once the range is determined, the top mounting hole location is calculated by
adding 4” to the upper measurement value. This is the distance above the
floor that the top mounting hole will be drilled into the wall.
Below are examples of possible measurement ranges and the required top
mounting hole.
TOP END CAP
MEASUREMENT RANGE
MOUNTING HOLE HEIGHT
39” to 90”
94”
35” to 86”
90”
30” to 81”
85”
25” to 76”
80”
Note: When measuring in centimeters, the maximum upper value is 230cm.
The top mounting hole is the upper value plus 10cm.
13
Measuring Height
1. Press and release the on/off button.
“Strt” will be flashing.
Reference
dot
2. Move the headpiece so it passes
the reference dot. “Strt” will stop
flashing and a beep will sound.
Raise the headpiece so it will be
well above the patient’s head.