background image

 

Selecting the Measurement Range 

The HTR-101 has a 51” measurement span with which you may select the 
upper and lower measurement values.  The only limitation is the upper value 
can not exceed 90” or 7ft 6in. 
 
Once the range is determined, the top mounting hole location is calculated by  
adding 4” to the upper measurement value.  This is the distance above the 
floor that the top mounting hole will be drilled into the wall. 
 
Below are examples of possible measurement ranges and the required top 
mounting hole. 
 
                                                                   
 

 

 

 

 

TOP END CAP 

MEASUREMENT RANGE 

   MOUNTING HOLE HEIGHT   

 
 

39” to 90” 

 

 

           94” 

 
 

35” to 86” 

 

 

           90” 

 
 

30” to 81” 

 

 

           85” 

 
             25” to 76” 

 

 

          80” 

 
 

 
 
Note:  When measuring in centimeters, the maximum upper value is 230cm.  
The top mounting hole is the upper value plus 10cm. 

 

13 

Measuring Height 

1.  Press and release the on/off button.  

“Strt” will be flashing. 

 
    

Reference 
dot 

2.  Move the headpiece so it passes 

the reference dot. “Strt” will stop 
flashing and a beep will sound. 

 
    Raise the headpiece so it will be 

well above the patient’s head. 

 
    

Summary of Contents for HTR-101

Page 1: ...2 Progress Drive Saukville WI 53080 Phone 262 284 5150 Toll Free 1 800 367 7109 Fax 262 284 5966 Email mail befour com www befour com 2014 Befour Inc Rev 1 Wall Mount Digital Height Rod INSTRUCTION MA...

Page 2: ...Number ___________________ Date Purchased ___________________ Call 800 367 7109 to obtain a Return Authorization number RA for warranty service Then send the product with freight and insurance to Bef...

Page 3: ...roduct is within the warranty period and the defect is in material or workmanship the product will be repaired or replaced and shipped back at no charge If the product is out of warranty or some other...

Page 4: ...ead height to minimize accidentally walking into the head piece Do not force the head piece slide past its upper and lower stops Always move the head piece slide only by the headpiece knob Do not use...

Page 5: ...inst the lower endcap is a constant You can record this value and quickly check whenever desired Why does the height rod display 5 7 3 as 5 7 3 This is a limitation of the bright LED digits We have fo...

Page 6: ...h display centimeter display feet inch display i e 79 3 i e 201 4 i e 6 7 3 5 To select your format release button when the desired format appears 6 Replace the battery cover Note Display format selec...

Page 7: ...ight measurement value displayed does not change while the headpiece is being raised For safety leave the headpiece in the top position when unit is not in use Keep magnets away from height rod You ma...

Page 8: ...ill be drilled into the wall Below are examples of possible measurement ranges and the required top mounting hole TOP END CAP MEASUREMENT RANGE MOUNTING HOLE HEIGHT 39 to 90 94 35 to 86 90 30 to 81 85...

Page 9: ...rom the guideline on the previous page determine the height of the top mounting hole and mark with a pencil Drilling a 1 8 pilot hole first to determine wall composition If you are mounting on drywall...

Page 10: ...e measure make a pencil mark on the side of the measurement rail 39 3 8 100 0 cm from the floor Check your mark to make sure it is correct 2 Turn on by pressing and releasing the on off button A flash...