41
Remove the Bung from the FeNObreath™.
Press the Home icon to return to the home
screen.
If the reading is not within the specification of the concentration range, one of the following should
be carried out:
1.
Calibrate using a CaliBag®
2.
Send the device back to MGC Diagnostics Corporation or local service center for
investigation
3.
Replace with a pre-calibrated sensor; MGC Diagnostics Corporation or the local service
center can do this for or a sensor replacement kit can be purchased separately which is
supplied with replacement instructions.
Calibration using a CaliBag®
NOTE:
The device must be calibrated annually at the following temperature of 21°C ±4°C (17°C-24°C)
or
70°F ±7°F (63°F-77°F)
The device will indicate whether the temperature range is out of specification by the following:
Device too cold to
be calibrated
Leave device in an atmospheric temperature of 21°C ±4°C (70°F
±7°F) until the temperature range within the device has
equilibrated.
Device too hot to
be calibrated
Leave device in an atmospheric temperature of 21°C ±4°C (70°F
±7°F) until the temperature range within the device has
equilibrated.